The FDA announced on Sept. 16, 2024, that Percussionaire Corporation is recalling Phasitron5 Breathing Circuits due to discolored fluid forming during therapy. The agency said the discoloration results from a reaction between medicated mist and an unapproved nickel coating on a spring component.
How to tell if you have it
- Brand: Percussionaire
- Product: Phasitron5 Breathing Circuit
- Model Numbers: P5-10, P5-HC, P5-UC-10
- Lot Codes: 240321, 240329, 240412, 240430, 240509, 240516, 240614, 240625, 240627
- Units Affected: 43,830 devices
The hazard
The FDA said breathing circuits are experiencing discolored fluid when therapy is given with specific medication through the nebulizer cup. This discoloration is caused by a reaction between medicated mist and an inadvertent or unapproved addition of a nickel coating on the spring component.
Context
GovInfoNow's records show Percussionaire Corporation has had seven other recalls since Sept. 14, 2020. Other recent recalls include an Intrapulmonary Percussive Ventilation labeling update (Aug. 18, 2026), a Phasitron 5 In-Line Valve risk of respiratory failure (Sept. 17, 2025), and a VDR4 Breathing Circuit venturi stuck risk (Dec. 23, 2024). Other recent recalls of similar products include Medline Medical Kits particulate in fluid path risks (March 24, 2026) and Hillrom VOLARA System P.CIRCUIT 5KIT AC oxygen leakage risk (May 21, 2026).