The FDA announced on Sept. 17, 2024, that Canon Medical System USA Inc. is voluntarily recalling its Cartesion Prime Digital PET-CT systems due to a hazard that may necessitate repeat examinations. The recall affects 27 units distributed across multiple U.S. states.
How to tell if you have it
- Brand: Canon Medical
- Product: Cartesion Prime, Digital PET-CT
- Model: PCD-1000A/3D.004, PCD-1000A/3D
- Software: V10.8 SP0013, V10.15 SP0006
- Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029
The hazard
According to the FDA, when the PET-CT system is executing reconstruction, if PET acquisition for another patient is performed or PET reconstruction for another patient is performed from raw data processing, PET reconstruction control processing may hang up and PET reconstruction may stop as a result. This malfunction may necessitate repeat PET/CT examinations.
Context
GovInfoNow's records show Canon Medical System USA Inc. has 26 other recalls on record since April 18, 2018. Other recent recalls of similar products include Canon Alphenix X-ray Systems Recall (Aug. 4, 2026), Canon Vantage Titan 3T MRI Systems Recall (Dec. 19, 2025), and Canon Alphenix INFX-8000H X-ray System Recall (Dec. 19, 2025).