Use caution Medical device recall · FDA · Published October 15, 2024

Draeger Atlan A300XL Anesthesia Workstation Recall: Risk of Ventilator Failure

Draeger recalls Atlan A300XL anesthesia workstations (model 8621400) due to potential piston ventilator failure before use or mechanical ventilation failure during use. Affected units have specific serial numbers.

🇺🇸 FDA recall #Z-0668-2025Medical deviceDraeger ›
Take action

Contact the company

Recalling firmDraeger Medical, Inc.3135 Quarry Rd · Telford, PA 18969-1042
✎ GovInfoNow plain-language explainer

What happened, in plain English

Draeger is recalling Atlan A300XL anesthesia workstations that could fail during use, risking patient safety. This recall affects medical facilities using the equipment for anesthesia procedures. Read more at govinfonow.com/recall/53704-draeger-atlan-a300xl-anesthesia-workstation-recall-risk-of-ventilator-failure.

Source Summary by GovInfoNow · govinfonow.com/recall/53704-draeger-atlan-a300xl-anesthesia-workstation-recall-risk-of-ventilator-failure · Updated August 18, 2026 · FDA enforcement record: FDA #Z-0668-2025

Cite this page

“Draeger Atlan A300XL Anesthesia Workstation Recall: Risk of Ventilator Failure.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/53704-draeger-atlan-a300xl-anesthesia-workstation-recall-risk-of-ventilator-failure. Based on FDA recall #Z-0668-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0668-2025%22&limit=1).

Why it’s dangerous

The anesthesia workstation may indicate a failure of the piston ventilator before use or suffer mechanical ventilation failure during use. This could lead to inadequate breathing support for patients during anesthesia procedures.

Who’s affected

Healthcare facilities using the Atlan A300XL anesthesia workstation for adult, pediatric, or neonatal procedures are most at risk. The recall applies to specific serial numbers shipped from Draegerwerk AG & Co., KGaA.

How to tell if you have this one

Check for the model number 8621400 on the Atlan A300XL anesthesia workstation. Affected units have specific serial numbers listed in the recall notice, including ASTB-0051 through ASTJ-0170.

  • Brand
    Draeger
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class I
  • Status
    Ongoing
  • Distribution
    No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
Recall numberProduct, codes & lotsQuantity
Z-0667-2025Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.; Model/Catalog Number: 8211300; UDI-DI Number: 04048675556343; Serial Numbers shipped directly from Draegerwerk AG & Co., KGaA to OUS: ASTA-0314; 0315; 0325; 0338; 0340; 0384; 0411; 0425 ASTB-0002; 0010; 0018; 0151; 0160; 0161; 0165; 0175; 0176; 0177; 0178; 0179; 0180; 0181; 0182; 0214; 0215; 0218; 0232; 0233; 0234; 0235; 0236; 0248; 0249; 0250; 0262; 0263; 0264; 0265; 0266; 0276; 0277; 0278; 0279; 0280 ASTC-0008; 0025; 0036; 0037; 0038; 0039; 0040; 0041; 0042; 0046; 0047; 0048; 0049; 0050; 0062; 0063; 0064; 0077; 0078; 0079; 0080; 0081; 0089; 0090; 0091; 0092; 0093; 0102; 0103; 0104; 0123; 0124; 0125; 0126; 0127; 0133; 0139; 0140; 0150; 0151; 0152; 0153; 0154; 0169; 0176; 0177; 0178; 0179; 0182; 0183; 0184; 0185; 0186; 0187; 0192; 0195; 0196; 0200; 0201; 0202; 0203; 0242 ASTD-0028; 0051; 0052; 0080; 0081; 0082; 0083; 0084; 0085; 0086; 0087 ASTE-0012; 0013; 0014; 0015; 0041; 0042; 0043; 0044; 0063; 0064; 0065; 0084; 0176; 0177; 0178; 0179; 0243; 0244; 0282; 0299 ASTF-0005; 0006; 0019; 0027; 0028; 0029; 0048; 0049; 0050; 0072; 0073; 0074; 0075; 0082; 0093; 0110; 0111; 0162; 0166; 0167; 0197; 0198; 0220; 0221; 0222; 0301; 0406; 0434 ASTH-0001; 0002; 0003; 0024; 0025; 0041; 0042; 0043; 0044; 0097; 0098; 0099; 0115; 0116; 0117; 0118; 0149; 0150; 0151; 0152; 0153; 0195; 0196; 0197; 0198; 0220; 0221; 0222; 0227; 0228; 0229; 0230; 0257; 0258; 0259; 0260; 0261; 0274; 0275; 0276; 0277; 0297; 0298; 0299; 0304; 0305; 0328; 0329; 0355; 0356; 0357; 0393; 0394; 0395; 0396; 0407; 0408; 0409; 0410; 0423; 0424; 0425; 0426; 0427; 0428; 0445; 0446; 0447; 0448; 0449; 0465; 0466; 0467; 0468; 0469; 0470; 0484; 0485; 0486; 0487 ASTK-0033; 0034; 0035; 0062280 units (OUS only)
Z-0668-2025Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.; Model/Catalog Number: 8621400; UDI-DI Number: 04048675556343; Serial Numbers shipped directly from Draegerwerk AG & Co., KGaA to OUS: ASTB-0051; 0053; 0054; 0055; 0063; 0065; 0093; 0162 ASTE-0219; 0220; 0221; 0222; 0377 ASTF-0472; 0473 ASTH-0005; 0006; 0026; 0027; 0047; 0186; 0223; 0224; 0401 ASTJ-017025 units (OUS Only)
Based on the official FDA remedy

What to do — step by step

  1. Check for Recalled Serial Numbers

    Review the serial numbers listed in the recall notice to determine if your Atlan A300XL is affected.

Where it went and when

Where it was distributed

Distributed in Colorado, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: yesCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFL

FDA distribution: “No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Colorado.

Recall timeline

FDA published this recall 2 months after the firm started it.

1. Firm started the recall: Oct 15, 202412. FDA classified it Class I: Dec 18, 202423. Published in FDA’s Enforcement Report: Dec 25, 202432024
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

What should I do if I have this product?

Check the serial numbers against the list provided in the recall notice to see if your Atlan A300XL is affected. No specific action is required if it is not on the list.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0668-2025 (October 15, 2024) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0668-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA