ICU Medical recalls counterfeit batteries used with Plum Infusion Systems that may fail to hold charge, causing early battery replacement messages. Affected units lack ICU Medical Test Label and CE Mark.
ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum Infusion Systems. Preliminary reports suggest that these non-OEM batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. OSI Batteries and their customers are distributing these non-OEM CSB batteries without authorization from ICU Medical.
ICU Medical recalls counterfeit batteries that may fail to hold charge and cause the infusion pump to display early battery replacement messages. These batteries are not tested for safety and could lead to pump malfunctions.
Counterfeit batteries from unauthorized sources may not hold charge properly, causing the infusion pump to show battery replacement messages prematurely. This could lead to pump malfunctions if the batteries fail unexpectedly.
Users of Plum 360" Infusion Systems and Plum A+ & A+3" Infusion Systems with counterfeit batteries that lack an ICU Medical Test Label and CE Mark. Patients using these systems are most at risk.
Check if your battery label lacks an ICU Medical Test Label and CE Mark. Counterfeit batteries have a date code (yellow label) in the range W2401xxxx to W2406xxx. Affected units include specific Plum Infusion System models listed in the recall.
Look for an ICU Medical Test Label and CE Mark on the battery. If missing, the battery is counterfeit and affected.
Counterfeit batteries may not hold charge properly, causing the infusion pump to show early battery replacement messages. This could lead to pump malfunctions if the batteries fail unexpectedly.
Counterfeit batteries lack an ICU Medical Test Label and CE Mark. They have a date code (yellow label) in the range W2401xxxx to W2406xxx.
Check if your battery lacks an ICU Medical Test Label and CE Mark. If so, stop using it and contact ICU Medical for guidance.
We’ve drafted a message you can send ICU Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0482-2025 — ICU Medical ICU Medical Hello, I own a ICU Medical that is covered by recall Z-0482-2025 from ICU Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.