Micro-X recalls specific mobile x-ray systems due to risks from improper temperature and humidity conditions causing capacitor issues. Affected models include MXU-RV35 with serial numbers 348-391.
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
Micro-X is recalling specific mobile x-ray systems (MXU-RV35) because they can malfunction if not used within strict temperature and humidity ranges. This could lead to overheating, unexpected startup, or capacitor failure if the device is operated outside the required environmental conditions.
The x-ray system requires precise temperature (14-131°F) and humidity (10-86%) levels to prevent condensation and ensure proper operation. If these conditions aren't met, capacitors may malfunction, causing overheating, unexpected startup, or other issues.
Healthcare facilities using the Micro-X Rover Mobile X-ray System # MXU-RV35 with serial numbers 348, 349, 361, 351, 365, or 391 are most at risk. These units were sold between 2021 and 2022.
Check for the model number MXU-RV35 on the device and serial numbers 348, 349, 361, 351, 365, or 391. These units were sold between 2021 and 2022.
Verify the current temperature and humidity levels to ensure they meet the required range of 14-131°F and 10-86% humidity.
The Micro-X Rover Mobile X-ray System can malfunction if operated outside specific temperature and humidity ranges, potentially causing overheating or unexpected startup.
Serial numbers 348, 349, 361, 351, 365, and 391.
Check for the model number MXU-RV35 and serial numbers 348, 349, 361, 351, 365, or 391. These units were sold between 2021 and 2022.
We’ve drafted a message you can send Micro-X to request your refund or repair — edit it as you like.
Subject: Recall Z-0580-2025 — Micro-X Micro-X Hello, I own a Micro-X that is covered by recall Z-0580-2025 from Micro-X. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.