Micro-X recalls two mobile x-ray systems (MXU-RV35 and MXU-RV71) due to potential under-reporting of post-exposure mAs values, which could affect radiation safety. Affected units have specific serial numbers and UDI-DI codes.
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Micro-X is recalling two mobile x-ray systems (MXU-RV35 and MXU-RV71) because they may report a lower mAs value than set by the operator. This could lead to incorrect radiation exposure measurements, potentially affecting patient safety.
The mobile x-ray system may report a lower post-exposure mAs value than what the operator sets. This could result in incorrect radiation exposure measurements, potentially leading to higher radiation doses for patients than intended.
Healthcare professionals using Micro-X Rover mobile x-ray systems with the specific serial numbers listed in the recall. Patients receiving scans on these units may be at risk if the mAs value is misreported.
Check for serial numbers MSN00334 through MSN00395 or UDI-DI codes 9357123000037 and 935712300005 on the Micro-X Rover Mobile X-ray System (models MXU-RV35 and MXU-RV71).
Verify if your unit has a serial number listed in the recall (MSN00334 to MSN00395).
Micro-X Rover Mobile X-ray System, models MXU-RV35 and MXU-RV71.
Check for serial numbers MSN00334 through MSN00395 or UDI-DI codes 9357123000037 and 935712300005 on the unit.
The recall record does not specify a remedy; the manufacturer is not providing a fix or action for affected units.
We’ve drafted a message you can send Micro-X to request your refund or repair — edit it as you like.
Subject: Recall Z-0665-2025 — Micro-X Micro-X Hello, I own a Micro-X that is covered by recall Z-0665-2025 from Micro-X. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.