Micro-X Ltd. recalls Rover Mobile X-ray Systems (model MXU-RV71) due to potential early x-ray tube failure from excessive thermal load. Affected units may require replacement to prevent safety risks.
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
Micro-X Ltd. is recalling their Rover Mobile X-ray System (model MXU-RV71) because the x-ray tube may fail early due to overheating. This could lead to unsafe imaging if the system is used without proper maintenance.
The x-ray tube in the Rover system can overheat during use, causing early failure. This might lead to improper imaging or system shutdowns, but it does not pose an immediate risk of injury.
Healthcare facilities using the Rover Mobile X-ray System (model MXU-RV71) are likely affected. Patients and medical staff using the system may be at risk if the x-ray tube fails during imaging.
Check for the model number MXU-RV71 on the system. Affected units have specific serial numbers starting with MSN0334 through MSN0427 as listed in the recall notice.
Look for the model number MXU-RV71 on the system to confirm if it's affected.
The recall is due to potential early x-ray tube failure from excessive thermal load, which could lead to improper imaging or system shutdowns.
Check for the model number MXU-RV71 and serial numbers starting with MSN0334 through MSN0427 as listed in the recall notice.
The recall does not specify a remedy, so contact Micro-X Ltd. directly for further instructions.
We’ve drafted a message you can send Rover to request your refund or repair — edit it as you like.
Subject: Recall Z-1578-2025 — Rover Rover Hello, I own a Rover that is covered by recall Z-1578-2025 from Rover. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.