The FDA announced on Nov. 25, 2024, that Spacelabs Healthcare, Inc. is recalling the Xhibit Telemetry Receiver, Model 96280, due to a hazard that may cause temporary loss of patient monitoring. The agency issued recall number Z-0809-2025 for the voluntary Class II action affecting 2,265 units distributed in the U.S. and internationally.
How to tell if you have it
- Brand: Spacelabs Healthcare
- Product: Xhibit Telemetry Receiver with Service Manual
- Model: 96280
- Service Manuals P/N: 070-2409-07 and Prior
- UDI-DI: 10841522107177, 10841522100246
The hazard
According to the FDA, telemetry receiver systems may experience a Windows proxy issue causing the system to shut down and restart for approximately 1 to 1.5 minutes. This results in an Offline display and temporary loss of monitoring, which could delay necessary clinical intervention if a clinical event occurred during that period.
Context
GovInfoNow's records show Spacelabs Healthcare, Inc. has 10 other recalls on record since Jan. 17, 2018. Recent brand recalls include the Spacelabs Xhibit Central Station Model 96102 (June 4, 2026) for audio/video display failures, the Spacelabs Ultraview SL Command Module Software Recall (March 26, 2026) for electrical leakage risk, and the Spacelabs Healthcare Multi-parameter Command Module Model 91496 Recall (Feb. 27, 2026) for inaccurate cardiac output readings. Other recent recalls of similar products include the GE HealthCare Telemetry Server Recall (April 24, 2026), the GE HealthCare CARESCAPE Telemetry Server V5 Recall (April 24, 2026), and the GE HealthCare CARESCAPE Telemetry Server ARK-2250L Recall (April 24, 2026).