Spacelabs Healthcare recalls Xhibit Telemetry Receiver Model 96280 due to temporary system shutdowns causing potential delays in detecting clinical events. Affected units may lose monitoring for 1-1.5 minutes during Windows proxy issues.
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
Spacelabs Healthcare is recalling Xhibit Telemetry Receiver Model 96280 units that may shut down temporarily, causing brief loss of patient monitoring. This could delay detection of critical clinical events if they occur during the shutdown period.
The system may shut down for 1-1.5 minutes due to a Windows proxy issue, causing temporary loss of patient monitoring. If a clinical event happens during this period, staff might not detect it until the system restarts, potentially delaying necessary medical intervention.
Healthcare facilities using the Xhibit Telemetry Receiver Model 96280 for patient monitoring are most at risk. Staff using these units in clinical settings may experience temporary monitoring loss during system shutdowns.
Check for the Xhibit Telemetry Receiver Model 96280 with Service Manual. Units sold with UDI-DI numbers 10841522107177 or 10841522100246 are affected.
Monitor for temporary shutdowns (1-1.5 minutes) causing loss of monitoring, which may indicate the issue.
The recall involves temporary system shutdowns that could delay detection of clinical events, but does not cause immediate harm or injury.
Affected units are the Xhibit Telemetry Receiver Model 96280 with Service Manual, sold with UDI-DI numbers 10841522107177 or 10841522100246.
The recall does not require immediate action, but monitor for temporary shutdowns that could delay clinical event detection.
We’ve drafted a message you can send Spacelabs Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0809-2025 — Spacelabs Healthcare Spacelabs Healthcare Hello, I own a Spacelabs Healthcare that is covered by recall Z-0809-2025 from Spacelabs Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.