The FDA announced on Dec. 12, 2024, that LivaNova USA, Inc. is recalling SenTiva and SenTiva Duo generators for the VNS Therapy System due to a hazard of stopping stimulation delivery. The agency issued recall number Z-1072-2025 for the voluntary Class II action.
How to tell if you have it
- Brand: LivaNova
- Product: SenTiva, REF: 1000
- Product: SenTiva Duo, REF: 1000-D
- Serial Numbers less than 500,000
The hazard
Vagus nerve stimulator generators may stop delivering stimulation because an internal, mechanically activated component can become stuck in a closed position. This failure may cause patients to return to baseline seizure frequency or depressive symptoms and potentially require generator replacement surgery.
Context
GovInfoNow's records show LivaNova USA, Inc. has 18 other recalls on record since Jan. 24, 2018. On Jan. 3, 2025, the company recalled implanted pulse generators due to a potential that they may stop delivering therapy because of an internal component issue. Other recent recalls include the Dec. 2, 2021, recall of VNS Therapy Leads Model 303 for risk of lead breach and the Nov. 12, 2020, recall of the SENTIVA DUO Device showing a wrong model number after reset.