The FDA announced on Jan. 15, 2025, that Getinge Usa Sales Inc is recalling the Vaporizer Sevoflurane, Quik-Fil (Model No. 6682285) due to a Class I hazard. The device may degrade sevoflurane into hydrogen fluoride, posing inhalation and skin exposure risks.
How to tell if you have it
- Brand: Getinge
- Product: Vaporizer Sevoflurane, Quik-Fil
- Model Number: 6682285
- UDI-DI: 07325710001141
- Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631
- Component of: Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e
- Quantity: 6 units
- Distribution: Domestic to Kentucky and Connecticut
The hazard
According to the FDA, sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of the respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Context
GovInfoNow's records show Getinge Usa Sales Inc has 16 other recalls on record since Dec. 9, 2021. Other recent recalls of similar products include Mindray Sevoflurane V90 Vaporizer Recall (Oct. 20, 2025), Mindray Isoflurane V90 Vaporizer Recall (Oct. 20, 2025), and Draeger Vapor 3000 Anesthetic Vaporizer Recall (Nov. 24, 2025).