Beckman Coulter Access 2 Immunoassay Analyzer Recall: Motor Alignment Issue May Delay Test

Beckman Coulter recalls Access 2 Immunoassay Analyzers with specific serial numbers due to a motor alignment issue that could delay patient test results. Affected units have a worn production tool causing sample motor misalignment.

🇺🇸 FDA recall #Z-1386-2025Medical deviceBeckman Coulter ›
Take action

Contact the company

Recalling firmBeckman Coulter, Inc.1000 Lake Hazeltine Dr · Chaska, MN 55318-1037
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling Access 2 Immunoassay Analyzers with certain serial numbers because a production tool issue caused sample motors to rub against the sample wheelbase plate. This could delay patient test results but does not cause immediate injury or damage. Read more at govinfonow.com/recall/52972-beckman-coulter-access-2-immunoassay-analyzer-recall-motor-alignment-issue.

Source Summary by GovInfoNow · govinfonow.com/recall/52972-beckman-coulter-access-2-immunoassay-analyzer-recall-motor-alignment-issue · Updated August 18, 2026 · FDA enforcement record: FDA #Z-1386-2025

Cite this page

“Beckman Coulter Access 2 Immunoassay Analyzer Recall: Motor Alignment Issue May Delay Test.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/52972-beckman-coulter-access-2-immunoassay-analyzer-recall-motor-alignment-issue. Based on FDA recall #Z-1386-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1386-2025%22&limit=1).

The FDA announced on Feb. 5, 2025, that Beckman Coulter, Inc. is recalling Access 2 Immunoassay Analyzers due to a potential delay in reporting patient test results. The recall affects units with specific serial numbers distributed worldwide.

How to tell if you have it

  • Brand: Beckman Coulter
  • Product: Access 2 Immunoassay Analyzer
  • Catalog Number: 81600N
  • UDI-DI: 15099590265380
  • Affected Component: Access 2 Sample Motors (B48016)
  • Units Affected: 873

The hazard

Beckman Coulter identified that improper alignment of the Access 2 Sample Motors causes them to rub against the sample wheelbase plate and index sensor. This issue, traced to a worn production tool at the supplier, could lead to a potential delay in reporting patient test results.

Context

GovInfoNow's records show Beckman Coulter has issued 132 other recalls since June 20, 2017. Recent recalls include the July 15, 2026, DxC 500 AU Modules recall for software defects and the June 26, 2026, MicroScan LabPro V5.03 Panel Update recall for incorrect algorithms. Other recent recalls of similar products include the Jan. 30, 2026, AU5800 Chemistry Analyzer recall and the Jan. 30, 2026, DxC 700 AU recall.

How to tell if you have this one

Check if your Access 2 Immunoassay Analyzer has the catalog number 81600N and the listed serial numbers (574626 through 574999, with specific numbers provided in the recall).

  • Brand
    Beckman Coulter
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Numbers

    Verify your Access 2 Immunoassay Analyzer's serial number against the list of affected numbers provided in the recall.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide - US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 49 days after the firm started it.

1. Firm started the recall: Feb 5, 202512. FDA classified it Class II: Mar 17, 202523. Published in FDA’s Enforcement Report: Mar 26, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Could this cause injury?

No injuries have been reported, and the hazard only affects test result timing, not physical safety.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1386-2025 (February 5, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1386-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA