Lower riskMedical device recall · FDA · Published March 3, 2025
Tornier Latitude EV Humeral Revision Stem Trial Recall: Incorrect Part Marking
Tornier recalls specific Latitude EV Humeral Revision Stem Trial models with incorrect part markings that could lead to improper fitting. Affected units have a laser marking of 9030103 but are actually 9030092. Consumers should contact Tornier for verification.
A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092, Humeral Stem Trial 150mm Medium / Left, Red color indicator (correct)
Take action
Contact the company
📍Recalling firmTornier, Inc10801 Nesbitt Ave S · Bloomington, MN 55437-3109
“Tornier Latitude EV Humeral Revision Stem Trial Recall: Incorrect Part Marking.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/52715-tornier-latitude-ev-humeral-revision-stem-trial-recall-incorrect-part. Based on FDA recall #Z-1651-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1651-2025%22&limit=1).
The FDA announced on March 3, 2025, that Tornier, Inc. is recalling one lot of stryker Latitude EV Humeral Revision Stem Trials due to incorrect direct part marking. The agency issued recall number Z-1651-2025 for the Class II device distributed internationally.
How to tell if you have it
stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103
UPC 00846832015054
Batch Number CC0421265
The hazard
According to the FDA, a specific lot of Latitude Humeral Trial Stems has incorrect direct part marking. The laser marking indicates REF 9030103 (Large/Right/Yellow), but the device is actually REF 9030092 (Medium/Left/Red).
Context
GovInfoNow's records show Tornier, Inc. has 12 other recalls on record since February 8, 2018. Other recent recalls of similar products include Encore Medical RSP Humeral Socket Insert Recall (January 15, 2026), restor3d Veritas Standard Glenoid Baseplate Recall (June 30, 2026), and W L Gore GORE VIABIL Short Wire Biliary Endoprosthesis Recall (May 22, 2026).
How to tell if you have this one
Check for the laser marking '9030103' on the device. The affected model is for the right side, large size, 150mm humeral revision stem trial. This model was sold with batch number CC0421265.
Brand
Tornier
Product
stryker Latitude EV HUMERAL REVISION STEM TRIAL, L: 150mm, Size: LARGE, Side: RIGHT, REF 9030103; Prosthesis, Elbow
Category
Medical device
Based on the official FDA remedy
What to do — step by step
Verify Device Marking
Compare the laser marking on the device to the correct model (9030092) to ensure it matches the intended size and side.
How this recall unfolded
Recall timeline
FDA published this recall 58 days after the firm started it.
1Firm started the recall
2FDA classified it Class II
3Published in FDA’s Enforcement Report
Dates from the agency record only; steps without a date are left out.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
Official source: FDA Recall #Z-1651-2025 (March 3, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1651-2025 to see it in context.
Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.