Maquet recalls VH-3010 Power Supply units that may cause HemoPro 2 harvesting tool to fail to heat properly due to incorrect resistor use. Affected units have specific serial numbers.
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Maquet is recalling its VH-3010 Power Supply units because they may have an incorrect resistor that prevents the HemoPro 2 harvesting tool from heating properly. This could lead to inadequate tissue processing during medical procedures.
The power supply uses an incorrect resistor that may prevent the HemoPro 2 harvesting tool from heating tissue adequately. This could result in incomplete tissue processing during medical procedures but does not pose an immediate risk of injury.
Medical professionals using the HemoPro 2 system with the recalled VH-3010 power supply. The risk is low as the hazard involves heating failure rather than immediate injury.
Check for the model number VH-3010 and serial numbers listed in the recall notice (e.g., H24080065G, H24080074G). The power supply is used with the Vasoview Hemopro EVH System (HemoPro 2).
Verify your power supply has model number VH-3010 and one of the listed serial numbers.
The power supply may have an incorrect resistor that causes the HemoPro 2 harvesting tool to fail to heat tissue properly during medical procedures.
Check if your power supply has model number VH-3010 and one of the listed serial numbers. If it does, contact Maquet for further instructions.
The recall indicates a risk of inadequate tissue heating during procedures, but it does not describe a risk of injury or serious harm.
We’ve drafted a message you can send Maquet to request your refund or repair — edit it as you like.
Subject: Recall Z-1577-2025 — Maquet Maquet Hello, I own a Maquet that is covered by recall Z-1577-2025 from Maquet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.