Thoratec HeartMate Mobile Power Unit Recall: Electrical Issues May Cause Unintended

Thoratec recalls HeartMate Mobile Power Units (MPUs) with serial numbers from 20226263 to 20307999. These units may experience electrical issues causing unintended shutdowns or alarms. If power is lost, the system may run for up to 15 minutes before stopping. Affected units were distributed between April 2024 and February 2025.

🇺🇸 FDA recall #Z-1570-2025Medical deviceThoratec ›
Take action

Contact the company

Recalling firmThoratec LLC6035 Stoneridge Dr · Pleasanton, CA 94588-3270
✎ GovInfoNow plain-language explainer

What happened, in plain English

Thoratec is recalling HeartMate Mobile Power Units (MPUs) that may have electrical issues causing unintended shutdowns or alarms. If the unit loses power, the system can run for up to 15 minutes before stopping. This could be dangerous for patients using left ventricular assist systems. Read more at govinfonow.com/recall/52635-thoratec-heartmate-mobile-power-unit-recall-electrical-issues-may-cause.

Source Summary by GovInfoNow · govinfonow.com/recall/52635-thoratec-heartmate-mobile-power-unit-recall-electrical-issues-may-cause · Updated August 18, 2026 · FDA enforcement record: FDA #Z-1570-2025

Cite this page

“Thoratec HeartMate Mobile Power Unit Recall: Electrical Issues May Cause Unintended.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/52635-thoratec-heartmate-mobile-power-unit-recall-electrical-issues-may-cause. Based on FDA recall #Z-1570-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1570-2025%22&limit=1).

Why it’s dangerous

The Mobile Power Unit may have an electrical component that causes it to shut down unexpectedly or restart without warning. This can trigger yellow wrench or no external power alarms. If the unit loses power, the system can support the pump for up to 15 minutes, but if not switched to rechargeable batteries, the pump may lose power.

Who’s affected

Patients using HeartMate 3 or II Left Ventricular Assist Systems with the recalled Mobile Power Units are most at risk. The recall affects units distributed between April 2024 and February 2025.

How to tell if you have this one

Check the serial number on the Mobile Power Unit. Affected units have serial numbers from 20226263 to 20307999. These units were distributed between April 2024 and February 2025 and are used with HeartMate 3 or II Left Ventricular Assist Systems.

  • Brand
    Thoratec
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Number

    Look for the serial number on the Mobile Power Unit to confirm if it's affected (20226263 to 20307999).

  2. Ensure Power Source

    If the unit loses power, the system can run for up to 15 minutes before stopping. Switch to rechargeable batteries if the unit is not powered by an external source.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 38 states.

Alaska: nationwideAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: named in the noticeORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: named in the noticeDEArizona: named in the noticeAZNew Mexico: named in the noticeNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: named in the noticeDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA, KS, AL, CO, ME and the countries of…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin.

Recall timeline

FDA published this recall 41 days after the firm started it.

1. Firm started the recall: Mar 13, 202512. FDA classified it Class I: Apr 17, 202523. Published in FDA’s Enforcement Report: Apr 23, 202532025
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1570-2025 (March 13, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1570-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA