Skip to content
Lower risk Medical device recall · FDA · Published April 23, 2025

Beckman Coulter DxI 9000 Access Analyzer Software Error Recall

Beckman Coulter recalls DxI 9000 Access Immunoassay Analyzers with software version 1.20 that may cause assay errors and delayed results due to incorrect motor/encoder installation.

🇺🇸 FDA recall #Z-1926-2025Medical deviceBeckman Coulter ›
Take action

Contact the company

Recalling firmBeckman Coulter, Inc.1000 Lake Hazeltine Dr · Chaska, MN 55318-1037
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling certain DxI 9000 Access Immunoassay Analyzers that have software version 1.20. These units may show error codes and delay results because of a faulty motor/encoder installation that requires field service to fix. Read more at govinfonow.com/recall/52426-beckman-coulter-dxi-9000-access-analyzer-software-error-recall.

Source Summary by GovInfoNow · govinfonow.com/recall/52426-beckman-coulter-dxi-9000-access-analyzer-software-error-recall · Updated August 18, 2026 · FDA enforcement record: FDA #Z-1926-2025

Cite this page

“Beckman Coulter DxI 9000 Access Analyzer Software Error Recall.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/52426-beckman-coulter-dxi-9000-access-analyzer-software-error-recall. Based on FDA recall #Z-1926-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1926-2025%22&limit=1).

Beckman Coulter, Inc. recalled DxI 9000 Access Immunoassay Analyzers on April 23, 2025, according to the FDA. The recall addresses incorrect motor/encoder installations that introduce error codes and delay test results. Owners should contact the company for resolution.

How to tell if you have it

  • Brand: Beckman Coulter
  • Product: DxI 9000 Access Immunoassay Analyzer
  • Catalog Number: C11137
  • Software versions: 1.20
  • Serial Numbers: 300144 through 300573 (specifically listed in record)
  • Quantity: 324 units

The hazard

The recall involves instruments manufactured starting with serial number 300141 where a new motor/encoder and cable were incorrectly installed. This error introduces codes for the DxI 9000 that prevent assays from being performed and delay result generation, potentially requiring field service.

Context

GovInfoNow's records show Beckman Coulter, Inc. has 132 other recalls on record since June 20, 2017. Other recent recalls of similar products include Beckman Coulter DxC 500 AU Modules with Software 1.5 Risk Patient Test Errors (July 15, 2026), Beckman Coulter MicroScan LabPro V5.03 Panel Update 11 DVD Kit: Incorrect Algorithms May Affect Test Results (June 26, 2026), and Beckman Coulter Access Total T4 Calibrator Recall: False Low Results Risk Incorrect Treatment (March 16, 2026). Additional similar recalls include AU480 Chemistry Analyzer May Show Error Message Causing Delay in Patient Results (Aug. 3, 2026), Beckman Coulter DxC 700 AU Recall: Delay in Results May Cause Diagnosis Delays (Jan. 30, 2026), and Beckman Coulter AU5800 Chemistry Analyzer Recall: Delay in Results Risk (Jan. 30, 2026).

How to tell if you have this one

Check if your DxI 9000 Access Immunoassay Analyzer has software version 1.20 and serial numbers listed in the recall (e.g., 300144, 300145, etc.).

  • Brand
    Beckman Coulter
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your analyzer's software version is 1.20 and check the serial number against the list provided in the recall.

  2. Contact Manufacturer

    Reach out to Beckman Coulter for field service assistance if your analyzer shows error codes or delayed results.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries of: Italy, Germany, Spain, New Zealand, France, Australia, Slovakia, Switzerland, Ireland, Brazil, Taiwan, Province of China, India, United Kingdom of Great…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 56 days after the firm started it.

1. Firm started the recall: Apr 23, 202512. FDA classified it Class II: Jun 6, 202523. Published in FDA’s Enforcement Report: Jun 18, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1926-2025 (April 23, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1926-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA