Lower risk Medical device recall · FDA · Published April 30, 2025

Beckman Coulter Sample Probe Recall: Defective Probe Mixes Sample Water

Beckman Coulter recalls sample probes with defective tips that mix cleaning water and sample, causing inaccurate test results. Affected units sold between specific dates and serial numbers.

🇺🇸 FDA recall #Z-1907-2025Medical deviceBeckman Coulter ›
Take action

Contact the company

Recalling firmBeckman Coulter Mishima K.K.454-32, Higashino; Nagaizumi-Cho · Sunto-Gun, Japan
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling sample probes that may mix cleaning water with the sample during use. This causes the sample to be diluted, leading to inaccurate test results. Affected units have specific serial numbers and were sold between certain dates. Read more at govinfonow.com/recall/52364-beckman-coulter-sample-probe-recall-defective-probe-mixes-sample-water.

Source Summary by GovInfoNow · govinfonow.com/recall/52364-beckman-coulter-sample-probe-recall-defective-probe-mixes-sample-water · Updated August 18, 2026 · FDA enforcement record: FDA #Z-1907-2025

Cite this page

“Beckman Coulter Sample Probe Recall: Defective Probe Mixes Sample Water.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/52364-beckman-coulter-sample-probe-recall-defective-probe-mixes-sample-water. Based on FDA recall #Z-1907-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1907-2025%22&limit=1).

Beckman Coulter Mishima K.K. recalled sample probes worldwide on April 30, 2025, according to the FDA. A manufacturing defect can cause cleaning water and samples to mix, leading to lower-than-expected analyte concentrations. The agency announced the Class II recall under number Z-1907-2025.

How to tell if you have it

  • Brand: Beckman Coulter
  • Product: Sample probe (S probe)
  • Lot numbers: 178713114 - 179433670
  • Device/Serial/Lot codes: AU480: B11810/14987666540664, B12183/14987666540671, B80091/16954701010015, B96692/14987666542965, B96693/14987666542972, C02654/14987666543283, C02655/14987666543290, C02845/16954701010022, C41919/14987666544174, C41920/14987666544181, N3659700/14987666535257, N3659800/14987666535264, N3659900/14987666535271, N3660000/14987666535288, N3660100/14987666535547, N3660200/14987666535554, N3660300/14987666535554, N3660400/14987666535769, N3660500/14987666536032, N3660600/14987666536049, N3660800/14987666536063, N3912400/14987666537084, N3912500/14987666537091, N3150700, N3151000, N3151300, N3151400, N3151500, N3151600, N3151900/
  • Serial numbers: From 10155 to 2025021482
  • Sample Probe codes: MU993400/04987666522939, Lots: From 178713114 to 179433670
  • B66751 Kits containing MU993400, Lots: From 178713114 to 179433670
  • Units affected: 10,176

The hazard

The FDA said a manufacturing defect in the sample probe causes the corners of the tip to be rounded and chamfered. This prevents the proper formation of an air layer that normally separates cleaning water from the sample. Consequently, the two mix and dilute the sample, resulting in analyte concentrations measured lower than expected.

Context

GovInfoNow's records show Beckman Coulter Mishima K.K. has had four other recalls on record since June 5, 2024. Other recent recalls of similar products include AU480 Chemistry Analyzer May Show Error Message Causing Delay in Patient Results (Aug. 3, 2026), Beckman Coulter DxC 700 AU Recall: Delay in Results May Cause Diagnosis Delays (Jan. 30, 2026), and Beckman Coulter DxC 500 AU Analyzer Recall: Software Error Causes Delayed Results (July 10, 2024). Additional recent recalls include Beckman Coulter CellMek SPS Sample Prep System Software Error Risk (Dec. 23, 2025), Beckman Coulter Bicarbonate Reagent May Cause False High Results - Recall (Oct. 14, 2025), and Beckman Coulter Access 2 Reaction Vessels Recall: Small Deformity May Cause Instrument Errors (Nov. 7, 2025).

How to tell if you have this one

Check the serial number on the probe or the lot number on the packaging. Affected probes have serial numbers from 10155 to 2025021482 and were sold with specific lots between 178713114 and 179433670.

  • Brand
    Beckman Coulter
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Completed
  • Distribution
    Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX, NJ, FL, IN, MA, NY, NC, OH, LA, AZ, MO, WA, GA, UT, KY, AL, CO, IL, NE, OK, NV, SC, TN, MD, OR, AR, MS, AK, WI, DE, MT, HI, KS, ME, NH, WV, NM, CT, IA, MI, MN, ID, RI, SD, WY, ND, DC, VT, GUAM, PR and OUS (foreign) to countries of: Viet Nam, Japan, India, Australia, El Salvador, China, United Arab Emirates, Thailand, Russian Federation, Italy, Germany, Korea (Republic of, Malaysia), Taiwan, Province of China, Turkey, Kazakhstan, Israel, Ecuador, Hungary, Egypt, Serbia, Bangladesh, Bosnia and Herzegovina, Costa Rica, Belarus, United Kingdom, Ireland, Botswana, Kenya, Mauritius, Ghana, Zambia, South Africa, Libya, Croatia, France, Switzerland, Czechia, Philippines, Namibia, Canada, Qatar, Azerbaijan, Lesotho, Bulgaria, Zimbabwe, Poland, Norway, Spain, Brazil, Tanzania (United Republic of), Lithuania, Yemen, Uzbekistan, Algeria, Somalia, Sri Lanka, Nepal, Pakistan, Ukraine, Peru, Morocco, Singapore, Slovakia, Mexico, Myanmar, Lebanon, Jordan, Malawi, Greece, Indonesia, Chile, Romania, Georgia, Ethiopia, Sweden, Albania, Eswatini, Nigeria, Cote d'Ivoire, Argentina, Dominican Republic, Iraq, Maldives, Saudi Arabia, Kuwait, Austria, Eritrea, Bolivia (Plurinational State of), Tunisia, Curacao, Mozambique, Hong Kong, Bahrain, Portugal, Panama, New Zealand, Colombia, Uganda, Oman, Belgium, Saint Pierre and Miquelon, Guyana, Macao, Cabo Verde, Sint Maarten (Dutch part), Liberia, Moldova, Republic of, North Macedonia, Congo (Democratic Republic of the), Uruguay, Cyprus, Netherlands, Suriname, Slovenia, Denmark, Finland, Venezuela, Sierra Leone
Recall numberProduct, codes & lotsQuantity
Z-1907-2025Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670; Device: REF/UDI-DI/Serial-Lot: AU480: B11810/14987666540664, B12183/14987666540671, B80091/16954701010015, B96692/14987666542965, B96693/14987666542972, C02654/14987666543283, C02655/14987666543290, C02845/16954701010022, C41919/14987666544174, C41920/14987666544181, N3659700/14987666535257, N3659800/14987666535264, N3659900/14987666535271, N3660000/14987666535288, N3660100/14987666535547, N3660200/14987666535554, N3660300/14987666535554, N3660400/14987666535769, N3660500/14987666536032, N3660600/14987666536049, N3660800/14987666536063, N3912400/14987666537084, N3912500/14987666537091, N3150700, N3151000, N3151300, N3151400, N3151500, N3151600, N3151900/ Serial: From 10155 to 2025021482. Sample Probe: MU993400/04987666522939, Lots: From 178713114 to 179433670. B66751 Kits containing MU993400, Lots: From 178713114 to 17943367010,176
Z-1908-2025Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670; Device: REF/UDI-DI/Serial-Lot: AU680: B12185/14987666540732, B12186/14987666540688, B12187/14987666540701, B12188/14987666540725, B96694/14987666542989, B96695/14987666542996, B96696/14987666543009, C02656/14987666543306, C02657/14987666543313, N3147700/15099590346249, N3662000/14987666535301, N3662100/14987666535318, N3662200/14987666535325, N3662700/14987666535615, N3663000/14987666535646, N3910200/14987666535790, N3910400/14987666535813, N3910500/14987666535820, N3910600/14987666535912, N3910800/14987666535936, N3910900/4987666535946, N3911200/14987666535974, N3911700/14987666536575, N3911900/14987666536599, N3147100, N3147200, N3147300, N3147400, N3147600, N3148100, N3149400, N3149800. Serial: From 10436 to 2022087656. Sample Probe: MU993400/04987666522939, Lots: From 178713114 to 179433670. B66752 Kits containing MU993400, Lots: From 178713114 to 1794336706,166
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Number

    Look for the serial number on the probe or the lot number on the packaging to determine if it's affected.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 50 states.

Alaska: named in the noticeAKMaine: named in the noticeMEVermont: named in the noticeVTNew Hampshire: named in the noticeNHWashington: named in the noticeWAIdaho: named in the noticeIDMontana: named in the noticeMTNorth Dakota: named in the noticeNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: named in the noticeRIMassachusetts: named in the noticeMAOregon: named in the noticeORNevada: named in the noticeNVWyoming: named in the noticeWYSouth Dakota: named in the noticeSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: named in the noticeWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: named in the noticeDEArizona: named in the noticeAZNew Mexico: named in the noticeNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: named in the noticeDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: named in the noticeHITexas: named in the noticeTXFlorida: named in the noticeFLPuerto Rico: named in the noticePR

FDA distribution: “Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX, NJ, FL, IN, MA, NY, NC, OH, LA, AZ, MO, WA, GA, UT, KY, AL, CO, IL, NE, OK, NV, SC, TN, MD, OR, AR, MS, AK, WI, DE, MT, HI, KS, ME, NH, WV, NM, CT, IA, MI, MN,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Apr 30, 202512. FDA classified it Class II: Jun 4, 202523. Published in FDA’s Enforcement Report: Jun 11, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1907-2025 (April 30, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1907-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA