Skip to content
Lower risk Medical device recall · FDA · Published June 17, 2025

Medtronic SmartSync Software Update Recalled: Erroneous Reset Message Risk

Medtronic recalls SmartSync Patient Connectors with specific software versions due to potential erroneous reset messages that could lead to unnecessary device explant. 39 reports of this error exist.

🇺🇸 FDA recall #Z-2127-2025Medical deviceMedtronic ›
Take action

Contact the company

Recalling firmMedtronic, Inc.8200 Coral Sea St Ne · Mounds View, MN 55112-4391
✎ GovInfoNow plain-language explainer

Medtronic is recalling SmartSync Patient Connectors with certain software versions because they may display a false reset message. This could cause doctors to unnecessarily remove the device, though no actual harm has occurred. The recall includes 39 reported cases of this issue. Read more at govinfonow.com/recall/52028-medtronic-smartsync-software-update-recalled-erroneous-reset-message-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/52028-medtronic-smartsync-software-update-recalled-erroneous-reset-message-risk · Updated September 23, 2026 · FDA enforcement record: FDA #Z-2127-2025

Cite this page

“Medtronic SmartSync Software Update Recalled: Erroneous Reset Message Risk.” GovInfoNow, updated September 23, 2026. https://govinfonow.com/recall/52028-medtronic-smartsync-software-update-recalled-erroneous-reset-message-risk. Based on FDA recall #Z-2127-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2127-2025%22&limit=1).

The FDA announced on June 17, 2025, that Medtronic, Inc. is recalling SmartLink Software loaded on SmartSync Patient Connector Model Number 24967. The software update eliminates the potential display of an erroneous electrical reset message during interrogation. Owners should obtain the available software update to correct this issue.

How to tell if you have it

  • Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967
  • UPC 00763000002039
  • UPC 00763000002046
  • UPC 00763000002053
  • UPC 00763000397852
  • UPC 00763000397869
  • UPC 00763000397876
  • UPC 00763000397883
  • UPC 00763000397890
  • UPC 00763000397906
  • UPC 00763000397913
  • UPC 00763000544300

The hazard

The erroneous message indicates an electrical reset has occurred, but no reset has actually happened and there are no parameter changes or loss of diagnostics. Prior to this software update, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through April 30, 2025.

Context

GovInfoNow's records show Medtronic, Inc. has 38 other recalls on record since September 10, 2017. Other recent recalls include the Medtronic Sphere-9 Catheter Recall (March 11, 2026), the Medtronic Abre Venous Stent Recall (December 4, 2025), and the Medtronic EVX 60 Series recall (October 29, 2025). Other recent recalls of similar products include the Medtronic Restore CP App Model A71100 Software Issue (November 20, 2025), the Philips Telemetry Monitor 5500 1.4 GHz Recall (June 5, 2026), and the GE Medical Systems Portrait Compute V1.2 Software Recall (July 20, 2026).

How to tell if you have this one

Check if your SmartSync Patient Connector has software model numbers D00U003 through D00U022 as listed in the recall notice. The device was sold with specific software versions that include the listed model numbers.

  • Brand
    Medtronic
  • Product
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide
Recall numberProduct, codes & lotsQuantity
Z-2127-2025Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.; 1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.3909 units
Z-2128-2025Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.; 1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.5418 units
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Review the software version on your device to see if it matches the listed versions in the recall notice.

How this recall unfolded

Recall timeline

FDA published this recall 36 days after the firm started it.

1. Firm started the recall: Jun 17, 202512. FDA classified it Class II: Jul 17, 202523. Published in FDA’s Enforcement Report: Jul 23, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Medtronic recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2127-2025 (June 17, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2127-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on September 23, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA