Use caution Medical device recall · FDA · Published June 23, 2025

HeartMate Mobile Power Unit AC Power Cord Recall: Misaligned V-Lock Risk

HeartMate recalls Mobile Power Unit AC Power Cord model 107760 due to misaligned V-Lock feature causing unintended disconnect. Affected patients should check if their device uses this model.

🇺🇸 FDA recall #Z-2290-2025Medical deviceHeartMate ›
Take action

Contact the company

Recalling firmThoratec LLC6035 Stoneridge Dr · Pleasanton, CA 94588-3270
✎ GovInfoNow plain-language explainer

HeartMate is recalling its Mobile Power Unit AC Power Cord model 107760 because the V-Lock feature may not lock properly, causing the cord to disconnect unexpectedly. This could lead to the device losing power during critical medical use, potentially causing harm to patients with heart failure. Read more at govinfonow.com/recall/51980-heartmate-mobile-power-unit-ac-power-cord-recall-misaligned-v.

Source Summary by GovInfoNow · govinfonow.com/recall/51980-heartmate-mobile-power-unit-ac-power-cord-recall-misaligned-v · Updated August 17, 2026 · FDA enforcement record: FDA #Z-2290-2025

Cite this page

“HeartMate Mobile Power Unit AC Power Cord Recall: Misaligned V-Lock Risk.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/51980-heartmate-mobile-power-unit-ac-power-cord-recall-misaligned-v. Based on FDA recall #Z-2290-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2290-2025%22&limit=1).

Thoratec LLC voluntarily recalled the HeartMate Mobile Power Unit AC Power Cord (Model 107760) due to a misaligned V-Lock feature that may cause unintended disconnect, the FDA said on June 23, 2025.

How to tell if you have it

  • Brand: HeartMate
  • Product Name: Mobile Power Unit, AC Power Cord
  • Model/Catalog Number: 107760
  • Lot Code: US Model No 107760
  • UDI-DI: 05415067038258

The hazard

The misaligned V-Lock feature on the Mobile Power Unit power cords prevents the locking mechanism from properly engaging with the Mobile Power Unit, which may cause unintended disconnect.

Context

Thoratec LLC has issued five other recalls since June 25, 2024, including a May 13, 2026 recall of HeartMate 11V Backup Batteries for failure risks and an Oct. 10, 2025 recall of the Thoratec CentriMag Blood Pump because it may not securely lock into the motor. Other recent recalls of similar products include a March 13, 2026 AVID Medical Halyard CARDIAC CATH PACK SAN23CARDQ Kit for loose connection risks and a May 21, 2026 Bard PowerPICC Catheters recall for defective lidocaine ampoules.

How to tell if you have this one

Check for the model number 107760 on the AC Power Cord label. The Mobile Power Unit is part of the HeartMate 3" Left Ventricular Assist System (LVAS) used for patients with advanced heart failure.

  • Brand
    HeartMate
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class I
  • Status
    Ongoing
  • Distribution
    Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, Japan.
Recall numberProduct, codes & lotsQuantity
Z-2290-2025Brand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. The HeartMate II Left Ventricular Assist System is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. The LVAS may be used in any of two configurations. First, line power may be utilized through the Power Module or the Mobile Power Unit (MPU) to run the LVAD indefinitely, convenient for sedentary or sleeping periods. Second, portable Battery power may be utilized for limited periods, convenient for active periods. The MPU AC power cord is distributed either as part of the complete MPU device package or as a standalone accessory. (Labeling - Attachments E1 to E6) Component: Yes, the AC Power Cord is a component of the Mobile Power Unit; Lot Code: US Model No 107760; UDI-DI 05415067038258, For Lot Numbers, see Attachment F.53
Z-2291-2025Brand Name: HeartMate Product Name: Mobile Power Unit Model/Catalog Number: 107754 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. The HeartMate II Left Ventricular Assist System is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. The LVAS may be used in any of two configurations. First, line power may be utilized through the Power Module or the Mobile Power Unit (MPU) to run the LVAD indefinitely, convenient for sedentary or sleeping periods. Second, portable Battery power may be utilized for limited periods, convenient for active periods. The MPU AC power cord is distributed either as part of the complete MPU device package or as a standalone accessory. Component: Yes, the MPU is a component of the HeartMate 3 and HeartMate II Left Ventricular Assist Systems.; Lot Code: US Model No 107754; UDI-DI 05415067038234 Serial Numbers, see Attachment F.604
Based on the official FDA remedy

What to do — step by step

  1. Check Power Cord Model

    Look for the model number 107760 on the AC Power Cord label to confirm if it is affected.

  2. Ensure Proper Connection

    If the cord is affected, make sure the V-Lock engages fully before using the device to prevent unintended disconnects.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 38 states.

Alaska: nationwideAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: named in the noticeWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: named in the noticeDEArizona: named in the noticeAZNew Mexico: named in the noticeNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: nationwideDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin.

Recall timeline

FDA published this recall 58 days after the firm started it.

1. Firm started the recall: Jun 23, 202512. FDA classified it Class I: Aug 13, 202523. Published in FDA’s Enforcement Report: Aug 20, 202532025
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2290-2025 (June 23, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2290-2025 to see it in context.