Skip to content
Stop using now Medical device recall · FDA · Published July 2, 2025

Stryker SmartPump Tourniquet Recall: Risk of Bleeding or Stroke During Use

Stryker recalls disposable tourniquets that may detach during use, causing bleeding, stroke, or death. Affected models include 5921-018-135 and others. Stop using immediately if you have these items.

🇺🇸 FDA recall #Z-2277-2025Medical deviceStryker ›
Take action

Contact the company

Recalling firmStryker Corporation1941 Stryker Way · Portage, MI 49002-9711
✎ GovInfoNow plain-language explainer

Stryker is recalling disposable tourniquets that could detach during use, potentially causing severe bleeding, stroke, or death. This risk is specific to certain medical procedures like IVRA (intravenous regional anesthesia). Read more at govinfonow.com/recall/51918-stryker-smartpump-tourniquet-recall-risk-of-bleeding-or-stroke-during.

Source Summary by GovInfoNow · govinfonow.com/recall/51918-stryker-smartpump-tourniquet-recall-risk-of-bleeding-or-stroke-during · Updated August 17, 2026 · FDA enforcement record: FDA #Z-2277-2025

Cite this page

“Stryker SmartPump Tourniquet Recall: Risk of Bleeding or Stroke During Use.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/51918-stryker-smartpump-tourniquet-recall-risk-of-bleeding-or-stroke-during. Based on FDA recall #Z-2277-2025 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2277-2025%22&limit=1).

The FDA announced on July 2, 2025, that Stryker Corporation is recalling Stryker SmartPump Tourniquets because the cuff flange may detach from the bladder. This defect poses risks of hemorrhage, stroke, or death during use. Owners should contact Stryker for a replacement.

How to tell if you have it

  • Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP
  • UPC 07613154599257
  • UPC 07613154599295
  • UPC 07613154599370

The hazard

The tourniquet cuff flange may become detached from the bladder, potentially causing a leak during procedures. This can result in unexpected blood loss, cerebral hypotension, cardiovascular collapse, or death. For IVRA procedures, patient toxicity may also occur due to rapid systemic absorption of anesthetics.

Context

GovInfoNow's records show Stryker Corporation has issued 34 other recalls since November 17, 2017. Recent Stryker recalls include a June 8, 2026, recall for AHTO Tube Sets due to leaking handpieces and a May 6, 2026, recall for Neptune Smart Docking Stations citing electrical shock risks. Other recent recalls of similar products include the March 12, 2026, Stryker Color Cuff 18"/24" Quick Connect Dual Port Cuffs Recall and the June 9, 2026, Zimmer Sterile Tourniquet Hoses Recall.

How to tell if you have this one

Check for model numbers 5921-018-135, 5921-018-235, or 5921-024-235 on the packaging. These tourniquets were sold with GTINs 07613154599257, 07613154599295, or 07613154599370.

  • Brand
    Stryker
  • Product
    Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Stop Using Immediately

    If you have the recalled tourniquets, stop using them right away to prevent potential bleeding or serious injury.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Jul 2, 202512. FDA classified it Class II: Aug 7, 202523. Published in FDA’s Enforcement Report: Aug 13, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2277-2025 (July 2, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2277-2025 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA