Boston Scientific Neuromodulation Corporation recalled Vercise Deep Brain Stimulation Systems Surgical Implant Manuals, Document Number 92328632-09, due to wire break hazards in associated devices. The Food and Drug Administration announced the recall on July 8, 2025.
How to tell if you have it
- Brand: Boston Scientific
- Product: Vercise Deep Brain Stimulation Systems Surgical Implant Manual
- Document Number: 92328632-09
- UDI-DI: 08714729985044, 08714729985051
- Revision: Rev A
The hazard
Wire breaks in rechargeable deep brain stimulation implantable pulse generators prevent successful stimulation. Users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms, and Bluetooth connectivity challenges.
Context
GovInfoNow's records show Boston Scientific Neuromodulation Corporation has 12 other recalls on record since April 24, 2019. Other recent recalls include the Boston Scientific INFINION CX Lead Kits Recall (June 17, 2026), the Boston Scientific CSK Electrodes Recall (April 29, 2026), and the Boston Scientific SC-4318 Clik X Anchor Recall (April 29, 2026). Similar recent recalls of other products include the Boston Scientific Rapid Refill System recall (June 30, 2026), the Hamilton Medical IntelliCuff Recall (May 22, 2026), and the Maquet Bubble Sensor Recall (January 9, 2026).