Skip to content
Lower risk Medical device recall · FDA · Published August 29, 2025

Fresenius Optiflux Dialyzer Recall: Press-on Caps May Not Secure Properly

Fresenius Medical Care recalls Optiflux High Flux E-beam Dialyzers with press-on caps that may not secure properly. Affected models were manufactured after September 6, 2023. No remedy specified in recall notice.

🇺🇸 FDA recall #Z-0041-2026Medical deviceFresenius Medical Care ›
Take action

Contact the company

Recalling firmFresenius Medical Care Holdings, Inc.920 Winter St Bld 920 · Waltham, MA 02451-1521
✎ GovInfoNow plain-language explainer

Fresenius Medical Care is recalling four specific models of Optiflux High Flux E-beam Dialyzers because the caps were changed to a press-on design that may not stay securely attached. This could lead to improper use during dialysis treatments if the caps become loose. Read more at govinfonow.com/recall/51518-fresenius-optiflux-dialyzer-recall-press-on-caps-may-not-secure.

Source Summary by GovInfoNow · govinfonow.com/recall/51518-fresenius-optiflux-dialyzer-recall-press-on-caps-may-not-secure · Updated August 17, 2026 · FDA enforcement record: FDA #Z-0041-2026

Cite this page

“Fresenius Optiflux Dialyzer Recall: Press-on Caps May Not Secure Properly.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/51518-fresenius-optiflux-dialyzer-recall-press-on-caps-may-not-secure. Based on FDA recall #Z-0041-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0041-2026%22&limit=1).

The FDA announced on Aug. 29, 2025, that Fresenius Medical Care Holdings Inc. is voluntarily recalling Optiflux High Flux E-beam Dialyzers to ensure proper use of the medical device. The company changed the caps from a threaded version to a press-on version requiring firm pressing to secure them.

How to tell if you have it

  • Brand: Fresenius Medical Care
  • Product: Optiflux High Flux E-beam Dialyzer
  • Model Numbers: 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM
  • UDI-DI: 840861100149, 840861100156, 840861100163, 840861100170
  • Lots manufactured after SEP/06/2023
  • Quantity: 172,494,732 units

The hazard

The recall addresses the change in cap design from a threaded version to a press-on version. Users must press the caps firmly onto the dialyzer to securely affix the cap and ensure proper use of the device.

Context

GovInfoNow's records show Fresenius Medical Care Holdings Inc. has 44 other recalls on record since April 14, 2014. Recent brand recalls include a May 12, 2026, notice regarding Bicarby Dialysate leaks and a May 11, 2026, recall of the 5008X Blood Tubing Set for online port errors. Other recent recalls of similar products include Vantive Healthcare's Jan. 6, 2026, Prismaflex M10 Dialyzer recall and Vantive US Healthcare's Jan. 6, 2026, Prismaflex HF1000 Set and ST100 Set recalls for dialyzer chamber dislodgement risks.

How to tell if you have this one

Check for model numbers: 0500316E, 0500318E, 0500320E, or 0500325E. These dialyzers were manufactured after September 6, 2023. The recall includes specific lot numbers and UDI-DI codes.

  • Brand
    Fresenius Medical Care
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Model Number

    Identify the model number on your dialyzer to see if it matches the recalled models: 0500316E, 0500318E, 0500320E, or 0500325E.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 47 days after the firm started it.

1. Firm started the recall: Aug 29, 202512. FDA classified it Class II: Oct 3, 202523. Published in FDA’s Enforcement Report: Oct 15, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Fresenius Medical Care recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0041-2026 (August 29, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0041-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA