The FDA announced on Sept. 16, 2025, that Spacelabs Healthcare is recalling the Sentinel V11 Cardiology Information Management System due to risks of patient data mix-ups. Owners should contact the company for a remedy.
How to tell if you have it
- Brand: Spacelabs
- Product Name: Sentinel V11
- Model/Catalog Number: 92810
- Software Version: 11.6.0
- UPC: 10841522125751
The hazard
According to the FDA, patient demographic fields may default to those of a previously viewed patient during admission. Additionally, test data intended for one patient may be saved under another's record on systems configured with Resting/Rhythm ECG functionality.
Context
GovInfoNow's records show Spacelabs Healthcare has 4 other recalls on record since 2017-10-11. Other recent recalls of similar products include GE Medical Systems ECG Systems Software Error Risks Patient Data Mix-Up (2026-04-21) and Nova Biomedical StatStrip Glucose Meter System Recall: Patient Data Mix-Up Risk (2026-07-29).