The FDA announced on Sept. 17, 2025, that Percussionaire Corporation is recalling Phasitron 5 IPV In-Line Valves because the pressure relief seal may leak when used with a ventilator. Owners should stop using the devices immediately.
How to tell if you have it
- Brand: Percussionaire
- Product: Phasitron 5, IPV In-Line Valve (Intrapulmonary Percussive Ventilation)
- REF: P5-TEE-20 (pack of 20) / P5-TEE (individual)
- Lots: 240326, 250116, 240418, 250324, 240610, 250616, 240620, 241121, 230612, 240826, WO04884, 241118, WO06020, 241203, WO05019, 250111, WO04827, WO04756
- Units affected: 59,160
The hazard
GovInfoNow's records show the leak may cause respiratory acidosis, hypoxemia, or respiratory failure. The hazard is particularly high for neonates, infants, or patients with target tidal volumes less than 50mL or weight less than 10kg due to unrecognized hypoventilation.
Context
GovInfoNow's records show Percussionaire Corporation has issued seven other recalls since Sept. 14, 2020. Other recent recalls of similar products include Medline Neonatal Breathing Circuits Risk of Sporadic Failure in Infants (July 13, 2026), Medline Hudson RCI Breathing Circuits Recall: Risk of Infant Respiratory Failure (July 13, 2026), and Medline Breathing Circuits Recall: Risk of Infant Breathing Failure (July 13, 2026).