Skip to content
Unrated Medical device recall · FDA · Published October 1, 2025

Siemens MAGNETOM Cima.X Upgrade: Ice Blockage Risk in Helium System

Siemens recalls MAGNETOM Cima.X Upgrade model 11689304 due to potential ice blockage in helium venting system, risking helium leaks in MRI rooms.

🇺🇸 FDA recall #Z-0431-2026Medical deviceSiemens ›
Take action

Contact the company

Recalling firmSiemens Medical Solutions USA, Inc40 Liberty Blvd · Malvern, PA 19355-1418
✎ GovInfoNow plain-language explainer

Siemens is recalling its MAGNETOM Cima.X Upgrade MRI systems with model number 11689304 because ice can block helium venting paths. This could cause pressure buildup and helium leaks in the scanning room if a quench occurs. Read more at govinfonow.com/recall/51142-siemens-magnetom-cima-x-upgrade-ice-blockage-risk-in-helium.

Source Summary by GovInfoNow · govinfonow.com/recall/51142-siemens-magnetom-cima-x-upgrade-ice-blockage-risk-in-helium · Updated August 17, 2026 · FDA enforcement record: FDA #Z-0431-2026

Cite this page

“Siemens MAGNETOM Cima.X Upgrade: Ice Blockage Risk in Helium System.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/51142-siemens-magnetom-cima-x-upgrade-ice-blockage-risk-in-helium. Based on FDA recall #Z-0431-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0431-2026%22&limit=1).

The FDA announced on Oct. 1, 2025, that Siemens Medical Solutions USA, Inc. is recalling the MAGNETOM Cima.X Upgrade due to a potential ice blockage in the venting system that could rupture the helium containment and cause a leak. This voluntary Class I recall affects one unit distributed to California.

How to tell if you have it

  • Brand: Siemens
  • Product: MAGNETOM Cima.X Upgrade
  • Model Number: 11689304
  • UPC: 04056869975245
  • Serial Number: 237304

The hazard

There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Context

GovInfoNow's records show Siemens Medical Solutions USA, Inc. has had 182 other recalls on record since June 6, 2017. Other recent recalls of similar products include Canon Vantage Titan 3T MRI Systems Recall (Dec. 19, 2025), Canon Medical Vantage Galan 3T MRI Recall (Dec. 19, 2025), and I.T.S. Hand Locking Plates System (Jan. 23, 2026).

How to tell if you have this one

Check for model number 11689304 on the device. The UDI is (01)04056869975245(21)237304 and serial number 237304.

  • Brand
    Siemens
  • Product
    MAGNETOM Cima.X Upgrade. Model Number: 11689304
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Model Number

    Look for model number 11689304 on the device.

Where it went and when

Where it was distributed

Distributed in California, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: yesCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFL

FDA distribution: “US distribution to California. International distribution to Australia, Germany, Great Britian.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: California.

Recall timeline

FDA published this recall 49 days after the firm started it.

1. Firm started the recall: Oct 1, 202512. FDA classified it Class I: Nov 12, 202523. Published in FDA’s Enforcement Report: Nov 19, 202532025
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Is there a risk of serious injury from this issue?

The recall record does not describe a risk of serious injury, so the danger level is unknown.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0431-2026 (October 1, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0431-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA