Lower risk Medical device recall · FDA · Published October 1, 2025

Beckman Coulter UniCel DxI 600 Photometric Analyzer May Delay Results Due to Defective

Beckman Coulter recalls UniCel DxI 600 photometric analyzers with specific part numbers and serials due to potential z-axis movement errors causing delays in test results.

🇺🇸 FDA recall #Z-0428-2026Medical deviceBeckman Coulter ›
Take action

Contact the company

Recalling firmBeckman Coulter, Inc.1000 Lake Hazeltine Dr · Chaska, MN 55318-1037
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling certain UniCel DxI 600 photometric chemistry analyzers because defective parts in the gantry assembly may cause the device to pause or not be ready, delaying test results. This could affect healthcare providers using the equipment for laboratory testing. Read more at govinfonow.com/recall/51130-beckman-coulter-unicel-dxi-600-photometric-analyzer-may-delay-results.

Source Summary by GovInfoNow · govinfonow.com/recall/51130-beckman-coulter-unicel-dxi-600-photometric-analyzer-may-delay-results · Updated August 18, 2026 · FDA enforcement record: FDA #Z-0428-2026

Cite this page

“Beckman Coulter UniCel DxI 600 Photometric Analyzer May Delay Results Due to Defective.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/51130-beckman-coulter-unicel-dxi-600-photometric-analyzer-may-delay-results. Based on FDA recall #Z-0428-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0428-2026%22&limit=1).

The FDA announced on Oct. 1, 2025, that Beckman Coulter, Inc. is recalling its UniCel DxI 600 photometric chemistry analyzers due to a defect causing result delays. The agency classified the recall as Class II.

How to tell if you have it

  • Beckman Coulter UniCel DxI 600
  • Part Numbers: A71460, A30260
  • UDI/DI 15099590369224 (PN A71460)
  • UDI/DI 15099590341602 (PN A30260)
  • Serial Numbers: 902515, 902529, 902528, 902534, 902526, 902527, 902523, 902522, 902524, 902505, 902578, 902579, 902545, 902530, 902531, 902581, 902504, 902514, 902525, 902535, 902566, 902516, 902518, 902532, 902555, 902558, 902564, 902519, 902521, 902559, 902511, 902549, 902554, 902547, 902582, 902583 (PN A71460); 902536, 902541, 902517, 902563, 902565, 902557, 902513, 902507, 902543, 902560, 902562, 902506, 902508, 902574, 902533, 902572, 902540, 902569, 902568, 902577, 902510, 902580, 902567, 902573, 902570, 902571, 902538, 902553, 902552, 902512, 902539, 902561, 902575, 902544, 902542, 902556, 902537, 902509, 902520, 902546, 902548, 902551, 902550, 902576 (PN A30260)

The hazard

An internal investigation determined that defective sub-components within the gantry assembly may cause a z-axis CAD movement motion error. This defect can cause the analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

Context

GovInfoNow's records show Beckman Coulter, Inc. has 132 other recalls on record since June 20, 2017. Recent similar recalls include the AU5800 Chemistry Analyzer recall (Jan. 30, 2026) and the DxC 700 AU recall (Jan. 30, 2026). Other recent recalls of similar products include the REMISOL Advance Software Issue (April 29, 2026), the Access Total T4 Calibrator Recall (March 16, 2026), the MicroScan LabPro V5.03 Panel Update (June 26, 2026), and the DxC 500 AU Modules recall (July 15, 2026).

How to tell if you have this one

Check for part numbers A71460 or A30260 on the device. The affected units have specific serial numbers listed in the recall notice, including many combinations starting with 9025.

  • Brand
    Beckman Coulter
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide
Recall numberProduct, codes & lotsQuantity
Z-0428-2026Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer; A71460, UDI/DI 15099590369224, Serial Numbers: 902515, 902529, 902528, 902534, 902526, 902527, 902523, 902522, 902524, 902505, 902578, 902579, 902545, 902530, 902531, 902581, 902504, 902514, 902525, 902535, 902566, 902516, 902518, 902532, 902555, 902558, 902564, 902519, 902521, 902559, 902511, 902549, 902554, 902547, 902582, 902583; A30260, UDI/DI 15099590341602, Serial Numbers: 902536, 902541, 902517, 902563, 902565, 902557, 902513, 902507, 902543, 902560, 902562, 902506, 902508, 902574, 902533, 902572, 902540, 902569, 902568, 902577, 902510, 902580, 902567, 902573, 902570, 902571, 902538, 902553, 902552, 902512, 902539, 902561, 902575, 902544, 902542, 902556, 902537, 902509, 902520, 902546, 902548, 902551, 902550, 902576.80 units
Z-0429-2026Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer; 973100, UDI/DI 15099590265373 (Chaska), Serial numbers: 610458, 610455, 610679, 610579, 610432, 610430, 610434, 610664, 610682, 610765, 610500, 610541, 610544, 610545, 610562, 610547, 610585, 610595, 610475, 610453, 610781, 610580, 610444, 610704, 610586, 610684, 610589, 610588, 610493, 610659, 610570, 610498, 610748, 610496, 610560, 610573, 610574, 610577, 610601, 610651, 610680, 610642, 610641, 610683, 610747, 610468, 610772, 610441, 610476, 610489, 610488, 610492, 610532, 610534, 610533, 610535, 610640, 610457, 610464, 610790, 610559, 610687, 610571, 610552, 610555, 610582, 610531, 610568, 610594, 610773, 610490, 610668, 610784, 610812, 610556, 610550, 610600, 610565, 610558, 610557, 610551, 610527, 610685, 610730, 610686, 610494, 610753, 610590, 610593, 610546, 610724, 610538, 610540, 610439, 610491, 610581, 610649, 610466, 610789, 610723, 610575, 610443, 610776, 610734, 610518, 610519, 610517, 610520, 610563, 610569, 610787, 610459, 610775, 610572, 610561, 610597, 610445, 610440, 610567, 610596, 610497, 610774, 610454, 610426, 610431, 610442, 610433, 610435, 610539, 610578, 610509, 610418, 610483, 610484, 610515, 610425, 610503, 610510, 610452, 610526, 610602, 610481, 610487, 610482, 610507, 610505, 610506, 610514, 610499, 610504, 610511, 610423, 610524, 610525, 610512, 610564, 610424, 610513, 610480, 610516, 610521, 610419, 610415, 610416, 610456, 610461, 610460, 610472, 610470, 610477, 610471, 610463, 610462, 610467, 610469, 610465, 610587, 610639, 610614, 610616, 610608, 610605, 610645, 610634, 610623, 610626, 610615, 610633, 610622, 610609, 610621, 610672, 610678, 610674, 610671, 610667, 610673, 610675, 610677, 610650, 610662, 610692, 610646, 610693, 610652, 610661, 610663, 610654, 610655, 610689, 610653, 610656, 610657, 610694, 610688, 610658, 610660, 610725, 610729, 610732, 610731, 610719, 610726, 610703, 610709, 610718, 610715, 610717, 610716, 610698, 610702, 610714, 610711, 610705, 610706, 610697, 610710, 610755, 610757, 610754, 610756, 610767, 610766, 610769, 610768, 610785, 610788, 610800, 610793, 610794, 610795, 610801, 610814, 610811, 610758, 610786, 610744, 610804, 610736, 610737, 610742, 610750, 610802, 610806, 610791, 610759, 610762, 610738, 610739, 610740, 610741, 610792, 610805, 610808, 610733, 610761, 610763, 610798, 610803, 610735, 610810, 610760, 610749, 610799, 610809, 610764, 610745, 610743, 610746, 610829, 610832, 610833, 610830, 610831, 610827, 610828, 610825, 610826, 610822, 610807, 610813, 610821. UDI/DI 16954701000412 (Suzhou), Serial Numbers: 630104, 630105, 630160, 630159, 630162, 630110, 630140, 630175, 630143, 630103, 630108, 630174, 630099, 630112, 630113, 630136, 630107, 630102, 630100, 630106, 630115, 630114, 630132, 630111, 630101, 630109, 630146, 630139, 630116, 630126, 630117, 630118, 630120, 630119, 630121, 630122, 630123, 630124, 630125, 630129, 630128, 630130, 630131, 630134, 630135, 630137, 630141, 630142, 630133, 630127, 630164, 630172, 630173, 630144, 630156, 630145, 630157, 630158, 630155, 630154, 630149, 630152, 630170, 630171, 630150, 630151, 630153, 630163, 630148, 630147, 630138, 630194, 630197, 630202, 630203, 630178, 630184, 630166, 630186, 630182, 630176, 630205, 630179, 630180, 630181, 630195, 630187, 630161, 630169, 630167, 630165, 630193, 630191, 630192, 630190, 630177, 630188, 630183, 630196, 630199, 630168, 630219, 630208, 630217, 630230, 630223, 630214, 630210, 630235, 630232, 630201, 630226, 630227, 630218, 630233, 630221, 630222, 630216, 630215, 630206, 630207, 630211, 630231, 630209, 630224, 630234, 630228, 630220, 630225, 630229, 630204, 630212, 630213, 630259, 630260, 630252, 630281, 630282, 630272, 630265, 630279, 630256, 630255, 630269, 630242, 630248, 630241, 630264, 630275, 630270, 630249, 630253, 630271, 630267, 630278, 630254, 630263, 630243, 630246, 630273, 630280, 630238, 630251, 630257, 630258, 630274, 630240, 630244, 630247, 630262, 630237, 630266, 630236, 630261, 630245, 630307, 630300, 630301, 630315, 630284, 630319, 630327, 630296, 630292, 630304, 630305, 630311, 630310, 630312, 630324, 630291, 630321, 630323, 630325, 630289, 630316, 630317, 630308, 630306, 630318, 630299, 630314, 630302, 630313, 630290, 630285, 630298, 630293, 630295, 630303, 630320, 630309, 630294, 630287, 630286, 630331, 630288, 630322, 630328, 630337, 630351, 630341, 630345, 630344, 630343, 630342, 630340, 630334, 630330, 630332, 630336, 630338, 630333, 630339, 630347, 630346, 630348, 630326, 630198, 630189, 630276, 630353, 630185, 630352, 630329, 630277, 630283, 630297, 630349, 630200, 630239.657 units
Based on the official FDA remedy

What to do — step by step

  1. Check Device Details

    Identify the part numbers and serial numbers on your UniCel DxI 600 analyzer to confirm if it is affected.

How this recall unfolded

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Oct 1, 202512. FDA classified it Class II: Nov 4, 202523. Published in FDA’s Enforcement Report: Nov 12, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using this device?

No action is required beyond checking if your device is affected; the recall does not require immediate discontinuation.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0428-2026 (October 1, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0428-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA