The FDA announced on Oct. 2, 2025, that Philips Medical Systems Nederland B.V. is voluntarily recalling Philips Azurion 7M20 systems with FlexArm ceiling-mounted systems due to potential motorized movement issues. The recall affects model number 722234.
How to tell if you have it
- Brand: Philips
- Product: Philips Azurion 7M20 systems with FlexArm ceiling-mounted system
- Model Number: 722234
- Quantity: 172 units
The hazard
According to the FDA, the motorized longitudinal movement of the FlexArm stand may be inconsistent and eventually become unavailable. This issue is caused by grease leakage from bearings or excessive anti-corrosion oil reducing friction between rails and the friction wheel.
Context
GovInfoNow's records show Philips Medical Systems Nederland B.V. has 58 other recalls on record since May 26, 2021. Other recent recalls of similar products include Canon Alphenix INFX-8000V X-ray System Recall (Dec. 19, 2025), Canon Alphenix INFX-8000C X-ray System Recall (Dec. 19, 2025) and Canon Alphenix INFX-8000H X-ray System Recall (Dec. 19, 2025).