Lower risk Medical device recall · FDA · Published October 9, 2025

HeartMate 3 Controller Recall: Potential Backup Battery Fault Alarm

Thoratec recalls HeartMate 3 Controller models due to potential corrosion in backup battery cable connector causing fault alarms. Affected units may need replacement of backup battery.

🇺🇸 FDA recall #Z-0506-2026Medical deviceHeartMate ›
Take action

Contact the company

Recalling firmThoratec LLC6035 Stoneridge Dr · Pleasanton, CA 94588-3270
✎ GovInfoNow plain-language explainer

Thoratec is recalling HeartMate 3 Controller devices because of potential corrosion in the backup battery cable connector that could trigger a fault alarm. This issue does not cause immediate harm but requires attention to prevent false alarms or system failures. Read more at govinfonow.com/recall/51076-heartmate-3-controller-recall-potential-backup-battery-fault-alarm.

Source Summary by GovInfoNow · govinfonow.com/recall/51076-heartmate-3-controller-recall-potential-backup-battery-fault-alarm · Updated August 18, 2026 · FDA enforcement record: FDA #Z-0506-2026

Cite this page

“HeartMate 3 Controller Recall: Potential Backup Battery Fault Alarm.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/51076-heartmate-3-controller-recall-potential-backup-battery-fault-alarm. Based on FDA recall #Z-0506-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0506-2026%22&limit=1).

Thoratec LLC voluntarily recalled the HeartMate 3 Controller (Standalone) due to potential corrosion issues with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm, the FDA said on Oct. 9, 2025.

How to tell if you have it

  • Brand: HeartMate
  • Product Name: HeartMate 3 Controller (Standalone)
  • Model/Catalog Number: 106531US, HeartMate 3; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller
  • Lot Code: All units of the following models are impacted: Model 106531US UDI: 05415067039446 Model 106531INT UDI: 05415067038982 Model 106531LF2 UDI: 05415067039422
  • Distribution: Worldwide including US states AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and countries Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan

The hazard

The recall is due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

Context

GovInfoNow's records show Thoratec LLC has 5 other recalls on record since 2024-06-25. Other recent recalls of similar products include HeartMate 11V Backup Battery Recall (2026-05-13), Thoratec CentriMag Blood Pump Recall (2025-10-10), and HeartMate Mobile Power Unit AC Power Cord Recall (2025-06-23). Other recent recalls of similar products include Ventec VOCSN+Pro Ventilator Recall (2026-05-08), Maquet Bubble Sensor Recall (2026-01-09), and AVID Medical Halyard CARDIAC CATH PACK SAN23CARDQ Kit Recall (2026-03-13).

How to tell if you have this one

Check for the HeartMate 3 Controller models: 106531US, 106531INT, or 106531LF2. The recall applies to all units with the listed model numbers and UDI codes.

  • Brand
    HeartMate
  • Category
    Medical device
FDA recall event · 4 products

All 4 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
Recall numberProduct, codes & lotsQuantity
Z-0505-2026Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. Component: No; Lot Code: All units of the following models are impacted: Model 106524US UDI: 00813024013297 Model 106524INT UDI: 0081302401171 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.29059
Z-0506-2026Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS; Lot Code: All units of the following models are impacted: Model 106531US UDI: 05415067039446 Model 106531INT UDI: 05415067038982 Model 106531LF2 UDI: 05415067039422 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.35376
Z-0507-2026Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, EU Software Version: N/A Product Description: The HeartMate II LVAS is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. Component: No; Lot Code: All units of the following models are impacted: Model 106015 UDI: 00813024011224 Model 106016 UDI: 00813024011231 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.248
Z-0508-2026Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II¿ LVAS Implant Kit; Lot Code: All units of the following models are impacted: Model 106017 UDI: 00813024011248 Model 106762 UDI: 00813024011286 Model 107801 UDI: 00813024011996 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.2424
Based on the official FDA remedy

What to do — step by step

  1. Check Controller Model

    Identify your HeartMate 3 Controller model number from the product label or documentation.

  2. Contact Manufacturer

    Reach out to Thoratec for guidance on whether your controller needs a backup battery replacement.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 41 states.

Alaska: named in the noticeAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: named in the noticeORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: named in the noticeWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: named in the noticeDEArizona: named in the noticeAZNew Mexico: named in the noticeNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: named in the noticeDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: named in the noticeHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin.

Recall timeline

FDA published this recall 48 days after the firm started it.

1. Firm started the recall: Oct 9, 202512. FDA classified it Class II: Nov 14, 202523. Published in FDA’s Enforcement Report: Nov 26, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0506-2026 (October 9, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0506-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA