The FDA announced on Oct. 27, 2025, that Philips North America is recalling Azurion R3.0 interventional fluoroscopic X-ray systems due to a missing safety component. The recall affects units shipped between July 2024 and June 2025.
How to tell if you have it
- Brand: Philips
- Product: Azurion R3.0, Commercial Name Azurion 7 M20
- Model Number: 722234
- Reference C&R: 2025-IGT-BST-016
- Units affected: 110
The hazard
The recalled systems were shipped without the Source-to-Skin Distance (SSD) Spacer. This component is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.