The FDA announced on Nov. 7, 2025, that Intuitive Surgical is recalling the da Vinci X, Xi, and da Vinci 5 surgical systems due to a software error allowing instruments with bent ball screws to remain in use. The agency issued recall number Z-1232-2026 for the Class II device.
How to tell if you have it
- Brand: Da Vinci
- Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System
- Model/Catalog Number: IS4000; IS4200; IS5000
- Affected Component: 380647-xx, USM
- Lot Code: UDI: 00886874114216
- FDA Medical Device Listing Number: D531324
- Affected Product List: UDI: 00886874114216 FDA Medical Device Listing Number: D295931
- Affected Product List: UDI: 00886574119747 FDA Medical Device Listing Number: D531324
The hazard
According to the FDA, a software implementation error allowed instrument arms that failed an important diagnostic test to remain in clinical use. This test was designed to detect Universal Surgical Manipulator insertion axis ball screws exhibiting symptoms of being bent and misaligned, which is a precursor to potential ball screw fatigue failure or fracture.
Context
GovInfoNow's records show Intuitive Surgical has issued 53 other recalls since June 1, 2017. Recent brand recalls include the April 24, 2026, Da Vinci 5 dV5 Tower NIR Handheld Camera Recall for unintentional activation and the April 22, 2026, Intuitive Surgical Da Vinci Surgical System Recall regarding E-brake pin risk. Other recent recalls of similar products include the Mazor X Robotic System Software Errors Risk Incorrect Spinal Surgery (Dec. 10, 2025), Think Surgical TMINI Tracking Arrays: Risk of Misalignment Due to Loose Knob (May 8, 2026), and Olympus HX-400U-30 Ligating Device Recall: Risk of Unintentional Loop Fixation (Oct. 30, 2025).