Skip to content
Lower risk Medical device recall · FDA · Published November 14, 2025

Olymp: ShockPulse-SE Lithotripsy System Generator Recall - Transducer Damage Risk

Olympus recalls ShockPulse-SE Lithotripsy System Generator (SPL-G) due to risk of transducer plug or receptacle damage causing device malfunctions. Affected units have all serial numbers.

🇺🇸 FDA recall #Z-1479-2026Medical deviceOlympus ›
Take action

Contact the company

Recalling firmOlympus Corporation of the Americas3500 Corporate Pkwy · Center Valley, PA 18034-8229
✎ GovInfoNow plain-language explainer

Olympus is recalling its ShockPulse-SE Lithotripsy System Generator (model SPL-G) because it can remain in a blinking state, potentially damaging the transducer plug or receptacle. This could lead to device malfunctions during medical procedures. Read more at govinfonow.com/recall/50651-olymp-shockpulse-se-lithotripsy-system-generator-recall-transducer-damage-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/50651-olymp-shockpulse-se-lithotripsy-system-generator-recall-transducer-damage-risk · Updated August 18, 2026 · FDA enforcement record: FDA #Z-1479-2026

Cite this page

“Olymp: ShockPulse-SE Lithotripsy System Generator Recall - Transducer Damage Risk.” GovInfoNow, updated August 18, 2026. https://govinfonow.com/recall/50651-olymp-shockpulse-se-lithotripsy-system-generator-recall-transducer-damage-risk. Based on FDA recall #Z-1479-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1479-2026%22&limit=1).

The FDA announced on Nov. 14, 2025, that Olympus Corporation of the Americas is recalling the ShockPulse-SE Lithotripsy System - Generator (Model SPL-G) due to a hazard where the device may remain in a blinking phase. The agency said this issue stems from potential damage to the transducer plug or generator receptacle.

How to tell if you have it

  • Brand: Olympus
  • Product Name: ShockPulse-SE Lithotripsy System - Generator
  • Model/Catalog Number: SPL-G
  • UDI: 00821925044203
  • Serial Numbers: All

The hazard

Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

Context

GovInfoNow's records show Olympus Corporation of the Americas has 133 other recalls on record since 2017-08-18. Other recent recalls include the Olympus Tripp-Lite P580-025 DisplayPort Cable (2026-08-27), the Olympus M3 Gold Autoclavable Foroblique Telescope (2026-08-24), and the Olympus BiCOAG Hemostasis Probe (2026-08-19). Other recent recalls of similar products include the Olympus OER-Pro Recall (2025-10-31), the GE Allia Moveo X-ray System Recall (2026-04-20), and the GE Medical Systems Allia IGS 3 Pulse X-Ray System Recall (2026-04-20).

How to tell if you have this one

Check for the model number SPL-G on the generator. All units have serial numbers and are affected.

  • Brand
    Olympus
  • Category
    Medical device
FDA recall event · 3 products

All 3 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
Recall numberProduct, codes & lotsQuantity
Z-1477-2026Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.; Model/Catalog Number: SPL-SR; UDI: 00821925043824; Serial Numbers: All;602 units
Z-1478-2026Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.; Model/Catalog Number: SPL-S; UDI: 00821925044197; Serial Numbers: All;1082 units
Z-1479-2026Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.; Model/Catalog Number: SPL-G; UDI: 00821925044203; Serial Numbers: All;1684 units
Based on the official FDA remedy

What to do — step by step

  1. Check Generator Model

    Look for the model number SPL-G on the generator to confirm it's affected.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 4 months after the firm started it.

1. Firm started the recall: Nov 14, 202512. FDA classified it Class II: Feb 27, 202623. Published in FDA’s Enforcement Report: Mar 11, 2026320252026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Is this recall dangerous?

The recall is for potential damage to the transducer plug or receptacle, which could cause device malfunctions. However, the hazard is not described as causing serious injury in normal use.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1479-2026 (November 14, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1479-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 18, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA