The FDA announced on Nov. 19, 2025, that Steris Corporation is voluntarily recalling the AMSCO 7052HP Washer/Disinfector due to a potential electrical arcing hazard. The agency stated the risk remains internal within the unit.
How to tell if you have it
- Brand: Steris
- Product: AMSCO 7052HP Washer/Disinfector
- Model/Catalog Number: 7052HP
- Lot Code (UDI): 00724995176587, 00724995176594, 00724995176570
- Serial Numbers: 3619625011, 3621625009, 3621625012, 3621925006, 3621925009, 3622425006, 3622725007, 3623425005, 3623825002, 3624525002
- Quantity: 10 units
- Distribution: US states (Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin) and Australia
The hazard
According to the FDA, a wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. The agency noted that electrical arcing remains internal within the unit.
Context
GovInfoNow's records show Steris Corporation has 29 other recalls on record since June 15, 2017. Recent similar product recalls include the Philips Allura Xper FD10C Medical System Recall (Dec. 15, 2025) and the Philips Allura Xper FD10F Medical System Recall (Dec. 15, 2025), both citing risks of electrical short-circuits.