The FDA announced on Nov. 20, 2025, that CareFusion 303, Inc. is recalling BD Pyxis MedStation ES devices due to a firmware update causing drawer failures that may delay access to medications. The recall affects approximately 53,669 units distributed worldwide.
How to tell if you have it
- BD Pyxis MedStation ES, REF: 323
- BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324
- BD Pyxis MedStation ES Tower, REF: 352
- BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00
- BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00
- BD Pyxis Pro MedStation Main, REF: 1155-00
- BD Pyxis CII Safe ES, REF: 1116-00
- Firmware Version: 1.48
The hazard
The FDA said the Automated Dispensing Cabinets received a Half Height CUBIE drawer firmware update that caused a Cubie Insert event, prompting software to generate a duplicate address. This error causes the drawer to fail, which may lead to an inability or delay in accessing stored items or patient-specific medications.
Context
GovInfoNow's records show CareFusion 303, Inc. has 64 other recalls on record since 2017-06-12. Other recent recalls of similar products include BD Pyxis MedStation ES REF 323 Medication Cabinet Recall (2026-04-30), CareFusion labeling recall (2026-02-11), and BD Pyxis ES Enterprise Server Software Update Issue Recall (2025-12-03). Other recent recalls of similar products include Medtronic Restore CP App Model A71100 Software Issue: Rare Therapy Disruption Risk (2025-11-20), Fresenius Medical Care 5008X CAREsystem Recall: Software Anomalies Risk Treatment Stop (2025-10-25), and GE Medical Systems Portrait Compute V1.2 Software Recall: Patient Monitoring Risk (2026-07-20).