Skip to content
Lower risk Medical device recall · FDA · Published November 24, 2025

ICU Medical LifeShield DLM Software v2.2: Precision Concentration Limits Issue

ICU Medical recalls LifeShield Drug Library Management software v2.2 due to concentration limit precision error. This may cause incorrect drug dosing if users adjust limits beyond manufacturer recommendations.

🇺🇸 FDA recall #Z-0966-2026Medical deviceICU Medical ›
Take action

Contact the company

Recalling firmICU Medical, Inc.600 N Field Dr · Lake Forest, IL 60045-4835
✎ GovInfoNow plain-language explainer

ICU Medical is recalling LifeShield Drug Library Management software version 2.2 because it only allows concentration limits to be set to one decimal place (0.1) instead of three (0.001). This could lead to incorrect drug dosing if users change limits to match manufacturer recommendations. Read more at govinfonow.com/recall/50612-icu-medical-lifeshield-dlm-software-v2-2-precision-concentration-limits.

Source Summary by GovInfoNow · govinfonow.com/recall/50612-icu-medical-lifeshield-dlm-software-v2-2-precision-concentration-limits · Updated August 17, 2026 · FDA enforcement record: FDA #Z-0966-2026

Cite this page

“ICU Medical LifeShield DLM Software v2.2: Precision Concentration Limits Issue.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/50612-icu-medical-lifeshield-dlm-software-v2-2-precision-concentration-limits. Based on FDA recall #Z-0966-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0966-2026%22&limit=1).

The FDA announced on November 24, 2025, that ICU Medical is recalling its LifeShield Drug Library Management (DLM) and LifeShield Infusion Safety Software Suite v2.2 due to a software issue that may cause incorrect medication delivery.

How to tell if you have it

  • LifeShield Drug Library Management (DLM), Product No. 17003-02
  • LifeShield Infusion Safety Software Suite v2.2, UDI-DI 00887709127227

The hazard

The software only allows Concentration Limits to be defined to one digit of precision past the decimal point instead of three digits. This limitation may require users to change limits from those recommended by the drug manufacturer, potentially resulting in over- or under-delivery of medication.

Context

GovInfoNow's records show ICU Medical has issued 60 other recalls since June 15, 2017. Other recent recalls include the CADD Yellow Medication Cassette Reservoir Recall on August 20, 2026, and the Hotline Fluid Warming Set Recall on June 24, 2026.

How to tell if you have this one

Check for LifeShield Drug Library Management (DLM) software version 2.2 with Product No. 17003-02. The software also has UDI-DI 00887709127227.

  • Brand
    ICU Medical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your software version by checking the product number or UDI-DI code on the software.

Where it went and when

Where it was distributed

Distributed nationwide; the notice names 10 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: nationwideILWisconsin: nationwideWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: nationwideINOhio: nationwideOHPennsylvania: named in the noticePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: named in the noticeKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: named in the noticeTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: nationwideTXFlorida: nationwideFL

FDA distribution: “US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Iowa, Kentucky, Louisiana, Maryland, Mississippi, New York, Pennsylvania, Tennessee, Virginia.

Recall timeline

FDA published this recall 37 days after the firm started it.

1. Firm started the recall: Nov 24, 202512. FDA classified it Class II: Dec 19, 202523. Published in FDA’s Enforcement Report: Dec 31, 202532025
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0966-2026 (November 24, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0966-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA