Philips recalls Allura Xper FD20 OR Table models with serial numbers 103, 10, 85, 125 due to potential coolant contact with electrical components if drip tray is not installed, risking system shutdown.
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Philips is recalling Allura Xper FD20 OR Table models with specific serial numbers because the drip tray under the cooling unit may not be properly installed. If the drip tray is missing or not correctly installed, coolant could contact electrical parts, causing short circuits and system shutdowns.
The drip tray beneath the cooling unit may not be installed correctly in some systems. If this happens, coolant liquid could come into contact with electrical components, causing short circuits that might shut down the system.
Medical facilities using Philips Allura Xper FD20 OR Table models with serial numbers 103, 10, 85, or 125 are most at risk. Patients and staff using these systems could face system shutdowns if the drip tray is not properly installed.
Check for serial numbers 103, 10, 85, or 125 on the Allura Xper FD20 OR Table. The model number is 722023, and the system was sold with specific UDI-DI and serial numbers.
Ensure the drip tray under the cooling unit is properly installed and documented as part of the field correction.
Philips identified that in some systems, the drip tray beneath the cooling unit was not installed under previous field corrections or documentation was inadequate, which could lead to coolant contacting electrical parts.
If the drip tray is missing or not correctly installed, coolant could contact electrical components, causing short circuits and system shutdowns.
Check for serial numbers 103, 10, 85, or 125 on your Allura Xper FD20 OR Table model with serial number 722023.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1072-2026 — Philips Philips Hello, I own a Philips that is covered by recall Z-1072-2026 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.