Philips recalls Allura Xper FD20 Biplane models with serial numbers starting with 296-488. If the drip tray under the cooling unit isn't installed, coolant could contact electrical parts, causing short circuits and system shutdowns.
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Philips is recalling all Allura Xper FD20 Biplane systems with model number 722008 due to a potential risk of coolant contacting electrical components if the drip tray isn't properly installed. This could cause short circuits and system shutdowns.
If the drip tray beneath the cooling unit isn't installed or properly documented, coolant liquid could contact electrical components. This might cause electrical short-circuits, leading to system shutdowns.
Medical professionals using Philips Allura Xper FD20 Biplane systems with model number 722008 are most at risk if the drip tray under the cooling unit isn't installed correctly.
Check for model number 722008 on the system. Serial numbers listed include 296, 245, 141, 260, 13, 410, 196, 476, 488, 290, 339, 474, 383, 530, 722008530, 223, 468, 63.
Ensure the drip tray beneath the cooling unit is properly installed and documented.
Philips identified that in some systems, the drip tray under the cooling unit wasn't installed or documented correctly, which could cause coolant to contact electrical parts and lead to short circuits.
All models with serial numbers starting with 296, 245, 141, 260, 13, 410, 196, 476, 488, 290, 339, 474, 383, 530, 722008530, 223, 468, 63.
No, but you must ensure the drip tray is installed and documented to prevent potential short circuits.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1067-2026 — Philips Philips Hello, I own a Philips that is covered by recall Z-1067-2026 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.