Philips recalls Allura Xper FD20 Biplane systems with model 722013 where drip tray under cooling unit may not be installed, risking electrical shorts and system shutdowns.
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Philips is recalling all Allura Xper FD20 Biplane systems with model number 722013. The recall is because the drip tray under the cooling unit might not be properly installed or documented, which could cause coolant to contact electrical parts and lead to system shutdowns.
If the drip tray beneath the cooling unit is not properly installed or documented, coolant liquid could contact electrical components. This might cause electrical short-circuits, triggering system fuses and leading to system shutdowns.
Medical facilities using the Allura Xper FD20 Biplane with model number 722013 are most at risk. This affects systems where the drip tray was not installed or documented during previous field corrections.
Check for model number 722013 on the Allura Xper FD20 Biplane system. Serial numbers listed include 601, 230, 370, 27, 320, 193, 60, 373, 560, 400, 549, 506, 381, 382, 132, 169, 233, 236, 243, 526, 447, 163, 403.
Verify that the drip tray under the cooling unit is installed and properly documented.
Philips identified that in some systems the drip tray beneath the cooling unit was not installed or documented, which could cause coolant to contact electrical components and lead to system shutdowns.
Model number 722013 for the Allura Xper FD20 Biplane.
Check for model number 722013 on your Allura Xper FD20 Biplane system. Serial numbers listed include 601, 230, 370, 27, 320, 193, 60, 373, 560, 400, 549, 506, 381, 382, 132, 169, 233, 236, 243, 526, 447, 163, 403.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1071-2026 — Philips Philips Hello, I own a Philips that is covered by recall Z-1071-2026 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.