Skip to content
Lower risk Medical device recall · FDA · Published December 23, 2025

Beckman Coulter CellMek SPS Sample Prep System Software Error Risk

Beckman Coulter recalls CellMek SPS Sample Preparation System with software error causing sample loss and potential incorrect diagnosis. Affected units have specific model and software version.

🇺🇸 FDA recall #Z-1235-2026Medical deviceBeckman Coulter ›
Take action

Contact the company

Recalling firmBeckman Coulter, Inc.11800 Sw 147th Ave · Miami, FL 33196-2500
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling its CellMek SPS Sample Preparation System due to a software error that can cause samples to be dispensed too close to tube bottoms. This may lead to fluid line dislodgement, sample loss, spillage, and incorrect patient diagnosis. Read more at govinfonow.com/recall/50360-beckman-coulter-cellmek-sps-sample-prep-system-software-error-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/50360-beckman-coulter-cellmek-sps-sample-prep-system-software-error-risk · Updated August 17, 2026 · FDA enforcement record: FDA #Z-1235-2026

Cite this page

“Beckman Coulter CellMek SPS Sample Prep System Software Error Risk.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/50360-beckman-coulter-cellmek-sps-sample-prep-system-software-error-risk. Based on FDA recall #Z-1235-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1235-2026%22&limit=1).

The FDA announced on 2025-12-23 that Beckman Coulter, Inc. is recalling the CellMek SPS Sample Preparation System due to a software error causing sample loss and potential incorrect diagnosis. Owners should contact the company for remedy details.

How to tell if you have it

  • Brand: Beckman Coulter
  • Product: CellMek SPS Sample Preparation System
  • REF: C44603
  • Software Version: 2.3.90.0
  • UDI-DI: 15099590750312

The hazard

A software error causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid lines and result in sample loss or spillage. This may lead to the unintentional removal of 60-100 microliters from the final sample, potentially causing incorrect diagnosis or patient management.

Context

GovInfoNow's records show Beckman Coulter, Inc. has 132 other recalls on record since 2017-06-20. Recent brand recalls include Beckman Coulter DxC 500i Modules with Software 1.5 (2026-07-15), Beckman Coulter MicroScan LabPro V5.03 Panel Update 11 (2026-06-26), and Beckman Coulter Access Total T4 Calibrator Recall (2026-03-16). Other recent recalls of similar products include Beckman Coulter REMISOL Advance Software Issue (2026-04-29), Beckman Coulter Emit 2000 Quinidine Assay Recall (2026-07-13), and Medline Namic Preceptor Manifold Recall (2026-03-24).

How to tell if you have this one

Check for the CellMek SPS Sample Preparation System with software version 2.3.90.0. The product has a UDI-DI number 15099590750312.

  • Brand
    Beckman Coulter
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version on your CellMek SPS system. If it is 2.3.90.0, it is affected.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 6 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: named in the noticeILWisconsin: nationwideWIMichigan: nationwideMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: named in the noticeORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: nationwideCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: named in the noticeMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: named in the noticeALGeorgia: nationwideGAHawaii: nationwideHITexas: nationwideTXFlorida: nationwideFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Illinois, Massachusetts, Missouri, Oregon, Washington.

Recall timeline

FDA published this recall 50 days after the firm started it.

1. Firm started the recall: Dec 23, 202512. FDA classified it Class II: Feb 2, 202623. Published in FDA’s Enforcement Report: Feb 11, 2026320252026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1235-2026 (December 23, 2025) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1235-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA