Use caution Medical device recall · FDA · Published January 9, 2026

FUJIFILM FDR Visionary Suite CH-200 X-ray Device Recall: Risk of Overheating and Smoke

FUJIFILM Healthcare recalls specific models of X-ray high-voltage generators (CH-200) due to risk of overheating and smoke after power cut. Affected units have unique serial numbers and versions.

🇺🇸 FDA recall #Z-1299-2026Medical deviceFUJIFILM Healthcare ›
Take action

Contact the company

Recalling firmFUJIFILM Healthcare Americas Corporation81 Hartwell Ave Ste 300 · Lexington, MA 02421-3160
✎ GovInfoNow plain-language explainer

FUJIFILM Healthcare is recalling certain X-ray high-voltage generators (FDR Visionary Suite CH-200) because internal electrical contacts can weld, causing overheating and potential smoke even after power is turned off. This poses a fire risk if the device is used improperly. Read more at govinfonow.com/recall/50176-fujifilm-fdr-visionary-suite-ch-200-x-ray-device-recall.

Source Summary by GovInfoNow · govinfonow.com/recall/50176-fujifilm-fdr-visionary-suite-ch-200-x-ray-device-recall · Updated August 17, 2026 · FDA enforcement record: FDA #Z-1299-2026

Cite this page

“FUJIFILM FDR Visionary Suite CH-200 X-ray Device Recall: Risk of Overheating and Smoke.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/50176-fujifilm-fdr-visionary-suite-ch-200-x-ray-device-recall. Based on FDA recall #Z-1299-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1299-2026%22&limit=1).

The FDA announced on Jan. 9, 2026, that FUJIFILM Healthcare Americas Corporation is voluntarily recalling the FDR Visionary Suite X-ray high-voltage generator due to a hazard of overheating and potential smoke. The agency issued recall number Z-1299-2026 for the Class II device.

How to tell if you have it

  • Brand: FUJIFILM Healthcare
  • Product: FDR Visionary Suite
  • Model Number: CH-200
  • Version: 566-16130-23, 566-16130-31, or 566-16130-33
  • Quantity: 16 units

The hazard

The contacts of the electromagnetic contactors inside the cabinet were found to be welded, which may allow current to continue supplying to the power supply circuit even when the device is turned off. This condition can cause the circuit to overheat and damage surrounding parts, and it may also cause smoke.

Context

GovInfoNow's records show FUJIFILM Healthcare Americas Corporation has 13 other recalls on record since Oct. 14, 2021. Other recent recalls of similar products include OXOS MC2 Portable X-Ray System: Defective Circuit Board Overheats (June 11, 2026) and BD Pyxis MedStation ES Tower 352 Medication Cabinet Recall: Fingerprint Scanner Overheats Risk (April 30, 2026).

How to tell if you have this one

Check for the model number CH-200 and specific versions: 566-16130-23, 566-16130-31, or 566-16130-33. Affected units have unique serial numbers listed in the recall notice.

  • Brand
    FUJIFILM Healthcare
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Details

    Identify the model number (CH-200) and version numbers (566-16130-23, 566-16130-31, or 566-16130-33) on your X-ray generator.

  2. Stop Using Immediately

    If your device matches the recalled model and version, stop using it immediately to prevent overheating or smoke.

Where it went and when

Where it was distributed

Distributed nationwide; the notice names 8 states.

Alaska: named in the noticeAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: named in the noticeNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: nationwideMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: named in the noticeOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: nationwideTXFlorida: named in the noticeFL

FDA distribution: “US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alaska, California, Connecticut, Florida, Illinois, New Hampshire, Ohio, Wisconsin.

Recall timeline

FDA published this recall 33 days after the firm started it.

1. Firm started the recall: Jan 9, 202612. FDA classified it Class II: Feb 5, 202623. Published in FDA’s Enforcement Report: Feb 11, 202632026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1299-2026 (January 9, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1299-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA