Skip to content
Lower risk Medical device recall · FDA · Published March 25, 2026

Heraeus PALAMIX uno: Mixing rod may come loose after 17 months of use

Heraeus recalls PALAMIX uno dental mixers with material number 66057893 that may have a mixing rod that comes loose after 17 months of use. Reduce shelf life from 3 years to 1.5 years.

🇺🇸 FDA recall #Z-1997-2026Medical deviceHeraeus ›
Take action

Contact the company

Recalling firmHeraeus Medical GmbH (Dental Division)Philipp-Reis-Str. 8 · Wehrheim, Germany
✎ GovInfoNow plain-language explainer

Heraeus is recalling PALAMIX uno dental mixers that could have a mixing rod that comes loose after 17 months of use. This risk is rare but could cause the mixer to malfunction during dental procedures. Read more at govinfonow.com/recall/49599-heraeus-palamix-uno-mixing-rod-may-come-loose-after-17.

Source Summary by GovInfoNow · govinfonow.com/recall/49599-heraeus-palamix-uno-mixing-rod-may-come-loose-after-17 · Updated August 17, 2026 · FDA enforcement record: FDA #Z-1997-2026

Cite this page

“Heraeus PALAMIX uno: Mixing rod may come loose after 17 months of use.” GovInfoNow, updated August 17, 2026. https://govinfonow.com/recall/49599-heraeus-palamix-uno-mixing-rod-may-come-loose-after-17. Based on FDA recall #Z-1997-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1997-2026%22&limit=1).

The FDA announced on March 25, 2026, that Heraeus Medical GmbH (Dental Division) is recalling Heraeus and PALAMIX uno products due to a potential mixing rod failure. The agency issued recall number Z-1997-2026 for the voluntary Class III action affecting worldwide distribution.

How to tell if you have it

  • Brand: Heraeus
  • Product: Heraeus, PALAMIX uno
  • Material Number: 66057893
  • UDI-DI: 04260102133587
  • Batch: all batches with numbers below 223060

The hazard

Long-term stability testing determined that in units older than 1.5 years (17 months), the mixing rod may come loose from its lock during the mixing process. The notification reduces the shelf life from 3 years to 1.5 years (17 months).

Context

GovInfoNow's records show Heraeus Medical GmbH (Dental Division) has two other recalls on record since December 9, 2025. Other recent recalls of similar products include PALACOS MV+G pro 40/80 Dental Cement Recall (2025-12-09), Heraeus PALACOS R pro 40 and 80 recall (2026-08-20), Stryker Hoffmann II Connecting Rod Micro 3 x 50mm recall (2026-07-01), Zimmer Disposable Mixing Bowls with Spatula Recall (2026-03-30), and Stryker Hoffmann II Connecting Rod Micro 3 x 120mm recall (2026-07-01).

How to tell if you have this one

Check for the material number 66057893 on the PALAMIX uno mixer. Units older than 17 months (1.5 years) with this material number are affected.

  • Brand
    Heraeus
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class III
  • Status
    Ongoing
  • Distribution
    Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Recall numberProduct, codes & lotsQuantity
Z-1997-2026Heraeus, PALAMIX uno. Material Number: 66057893.; Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;51,771 units
Z-1998-2026Heraeus, PALAMIX duo. Material Number: 66057897.; Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;1,125 units
Based on the official FDA remedy

What to do — step by step

  1. Check Product Date

    Verify the material number is 66057893 and the mixer has been in use for more than 17 months.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Mar 25, 202612. FDA classified it Class III: Apr 29, 202623. Published in FDA’s Enforcement Report: May 6, 202632026
  1. Firm started the recall
  2. FDA classified it Class III
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1997-2026 (March 25, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1997-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA