The FDA announced on April 22, 2026, that Intuitive Surgical, Inc. is recalling Da Vinci ASSY, PSS, SP1098 components due to a potential E-brake retainer pin issue that may cause uncontrolled motion and patient injury. The agency classified the Class II recall as voluntary.
How to tell if you have it
- Brand: Intuitive Surgical
- Product: Da Vinci ASSY, PSS, SP1098
- Part Number: 380601
- UDI: 00886874114605
- Units affected: 454
The hazard
The operative components may be distributed with an E-brake retainer pin that prevents E-brake engagement during a structural column failure, potentially resulting in uncontrolled motion. If the pin moves or backs out, it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury, crushing injury, or bleeding.
Context
GovInfoNow's records show Intuitive Surgical, Inc. has 53 other recalls on record since June 1, 2017. Other recent recalls of similar products include Da Vinci 5 dV5 Tower NIR Handheld Camera Recall (April 24, 2026), Da Vinci X and Xi Surgical Systems Recall (April 2, 2026), and Intuitive 8 mm SureForm 30 Gray Reloads Recall (March 11, 2026). Other recent recalls of similar products include Vortex Surgical TID Pharos Illuminated Depressor Recall (December 16, 2025), Synthes 2.4V TAP CORTEX SCREW Recall (February 25, 2026), and Olympus HX-400U-30 Ligating Device Recall (October 30, 2025).