Bayer Medical Care, Inc. recalled Avanta Multi-Patient Administration Tube Sets on April 24, 2026, due to a fit issue that may cause the stopcock to disengage during use, according to the FDA. The agency announced the Class II recall of 11,300 units distributed worldwide.
How to tell if you have it
- Brand: Bayer
- Product: Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2))
- Catalog number: AVA 500 MPAT (Part 2 of 2)
- Affected material numbers: 87629007, 60729458, 86566648
- Affected batch numbers: 251702, 252302, 252502, 252602, 252702, 252802, 253304, 253702, 253704, 253804, 254202, 254304, 254704, 254802
The hazard
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface, and if forced, the connection is unstable, potentially leading to procedure interruption.
Context
GovInfoNow's records show Bayer Medical Care, Inc. has had 7 other recalls since June 28, 2018. Recent brand recalls include a July 31, 2026 sterilization issue with Stellant FLEX Syringe Kits and an Nov. 30, 2022 labeling error for Source Administration Sets. Other recent recalls of similar products include a Nov. 26, 2025 Medline kit recall involving stuck check valves and an April 3, 2026 Elekta applicator set recall regarding tube detachment risk.