The FDA announced on May 21, 2026, that Laborie Medical Technologies Corp is recalling the Solar GI System (Water Perfusion) due to potential moisture accumulation affecting performance. The agency issued recall number Z-2669-2026 for the voluntary Class II recall.
How to tell if you have it
- Brand: Laborie Medical
- Product Name: Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14)
- Catalog Numbers: G3-6, G3-7, G3-8, G3-12, G3-14
- Primary UDI-DI: 08717775955047, 08717775955054, 08717775955061, 08717775955078, 08717775955085
- Basic UDI-DI: 87177759500G3G8
- Quantity: 372 units
The hazard
According to the FDA, moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors during extended clinical use. This accumulation may affect airflow or pressure measurement performance if not properly removed.
Context
GovInfoNow's records show Laborie Medical Technologies Corp has one other recall on record since July 17, 2019: the Solar Blue Laptop Power Supply Recall for fire hazard risk. Other recent recalls of similar products include Olympus UHI-3 Insufflation Unit Recall (Jan. 16, 2026), Medline medical procedure kits with Strykeflow 2 irrigation leak risk (Dec. 16, 2025), and Olympus UHI-2 Insufflation Unit Recall (Jan. 16, 2026).