Hillrom recalls Volara System circuit kits with potential air leaks causing oxygen drops. Affected units have model REF M08473 and were sold starting April 28, 2025. Stop using if leaks occur.

🇺🇸 FDA recall #Z-2550-2026Medical deviceHillrom ›
Take action

Contact the company

Recalling firmBaxter Healthcare Corporation1 Baxter Pkwy · Deerfield, IL 60015-4625
✎ GovInfoNow plain-language explainer

Baxter Healthcare Corporation recalls Hillrom Volara System circuit kits that may leak air during therapy, risking low oxygen levels and ineffective medication delivery. This happens when the nebulizer cup isn't properly locked after adding medication. Read more at govinfonow.com/recall/49094-hillrom-volara-system-circuit-kit-air-leaks-risk-oxygen-drop.

Source Summary by GovInfoNow · govinfonow.com/recall/49094-hillrom-volara-system-circuit-kit-air-leaks-risk-oxygen-drop · Updated August 10, 2026 · FDA enforcement record: FDA #Z-2550-2026

Cite this page

“Hillrom Volara System Circuit Kit: Air Leaks Risk Oxygen Drop During Therapy.” GovInfoNow, updated August 10, 2026. https://govinfonow.com/recall/49094-hillrom-volara-system-circuit-kit-air-leaks-risk-oxygen-drop. Based on FDA recall #Z-2550-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2550-2026%22&limit=1).

Baxter Healthcare Corporation recalled Hillrom VOLARA System P.CIRCUIT KIT AC nebulizer circuits due to air and medication leakage risks, the FDA announced on May 21, 2026. The issue may cause oxygen level drops in patients.

How to tell if you have it

  • Brand: Hillrom
  • Product: VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit)
  • Model/Catalog Number: REF M08473
  • Lot Numbers: All lot numbers distributed beginning on 4/28/2025
  • Units affected: 16,187

The hazard

Reports indicate air and medication leakage from the nebulizer cup during therapy, potentially leading to oxygen level drops and ineffective nebulization. The FDA stated this issue is associated with improper locking of the nebulizer cup after adding medication. No specific incidents or injuries were reported in the record.

Context

GovInfoNow's records show Baxter Healthcare Corporation has had 187 other recalls since September 2, 2015. Recent Baxter recalls include a PRO+ 40" MRS SURFACE Mattress recall on August 18, 2026, for incorrect labeling risk and a Sodium Chloride Injection recall on August 17, 2026, for fiberglass particulates risk. Other recent recalls of similar products include NOXBOX Nitric Oxide Delivery System Recall (August 19, 2026), ICU Medical CADD Yellow Medication Cassette Reservoir Recall (August 20, 2026), and Hamilton Medical Breathing Circuit Sets Risk (May 29, 2026).

How to tell if you have this one

Check for the model number REF M08473 on the circuit kit. The product was sold starting April 28, 2025, with all lot numbers from that date onward.

  • Brand
    Hillrom
  • Category
    Medical device
FDA recall event · 4 products

All 4 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class I
  • Status
    Ongoing
  • Distribution
    Worldwide distribution - US Nationwide and the country of Canada.
Recall numberProduct, codes & lotsQuantity
Z-2548-2026Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device; UDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/202510540 units
Z-2549-2026Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device; UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/20250
Z-2550-2026Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device; UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/202516187 units
Z-2551-2026Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device; UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/202513881 units
Based on the official FDA remedy

What to do — step by step

  1. Check for Leaks

    Look for air leaks around the nebulizer cup after adding medication. If leaks occur, stop using the circuit kit immediately.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide and the country of Canada.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 48 days after the firm started it.

1. Firm started the recall: May 21, 202612. FDA classified it Class I: Jul 1, 202623. Published in FDA’s Enforcement Report: Jul 8, 202632026
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Hillrom recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2550-2026 (May 21, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2550-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 10, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA