Tosoh Bioscience Inc recalled ST AIA-Pack Testosterone in vitro diagnostic kits on June 12, 2026, according to the FDA. The recall affects devices that may produce erroneous testosterone results due to low-quality control values.
How to tell if you have it
- Brand: Tosoh
- Product: ST AIA-Pack Testosterone
- Model Number: 0025204
- UDI-DI: 14560189210032
- Lot Numbers: FX12404, FY12407, FY12406, FZ12408
- Units affected: 310
The hazard
The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA), which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.
Context
GovInfoNow's records show Tosoh Bioscience Inc has 19 other recalls on record since 2017-06-11. Other recent recalls of similar products include Siemens Atellica CH Urine Albumin Recall (2026-07-06), Immunotech Estrone RIA Kit Recall (2025-12-15), and Vivoo Test Strips Recall (2025-11-22).