The FDA announced on June 16, 2026, that Accuray Incorporated is recalling Cyberknife Treatment Delivery Systems equipped with the Xchange Robotic Collimator Changer. The hazard involves a collimator potentially falling from its housing if it fails to fully dock and personnel use the in-room control pendant.
How to tell if you have it
- Brand: Accuray
- Product: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
- Software: Version 11.X and higher that utilize the Exchange Table
- Quantity: 292 units
The hazard
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Context
GovInfoNow's records show Accuray Incorporated has 9 other recalls on record since November 17, 2017. Other recent recalls of similar products include Canon Alphenix X-ray Systems Recall: Loose Fixing Screws Risk Equipment Failure (August 4, 2026), Canon Alphenix INFX-8000C X-ray System Recall: Ceiling Movement May Become Unstable (December 19, 2025), and Merit Medical Custom Procedure Kit Recall: Handle May Detach During Use (December 4, 2025).