The FDA announced on May 21, 2026, that Instrumentation Laboratory is recalling GEM Premier 5000 PAKs due to lysing bag solution concentration errors. This malfunction could delay the recognition of Co-Oximetry and total bilirubin abnormalities. Owners should contact the company for a remedy.
How to tell if you have it
- Brand: Instrumentation Laboratory
- Product: GEM Premier 5000 PAK
- Part Numbers: 00055360010, 00055360011, 00055407504, 00055415010, 00055430010, 00055430011, 00055445010
- Part No. 00055360010 UDI-DI 08426950807599 Lots 260505K, 260504AG, 260505F
- Part No. 00055360011 UDI-DI 08426950807605 Lots 260505F and 260505J
- Part No. 00055407504 UDI-DI 08426950807537 Lot 260505R
- Part No. 00055415010 UDI-DI 08426950807698 Lots 260505H and 260505K
- Part No. 00055430010 UDI-DI 08426950807759 Lots 260505H, 260505K and 260505F
- Part No. 00055430011 UDI-DI 08426950807766 Lot 260505K
- Product No. 00055445010 UDI-DI 08426950807810 Lots 260504AA and 260505K
- Product No. 00055445011 UDI-DI 08426950807827 Lots 260505F and 260505J
- Product No. 00055360010.A UDI-DI 08426950807599 Lot 260504AF
The hazard
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Context
GovInfoNow's records show Instrumentation Laboratory has 13 other recalls on record since 2021-08-25. Other recent recalls of similar products include GEM Premier 5000 Blood Analyzer (2026-05-08), GEM Premier 5000 PAK Cartridges (2026-01-20), and Instrumentation Laboratory Coagulation Test Kits (2025-12-11).