The FDA announced on August 18, 2026, that Radiometer Medical ApS is recalling The ABL90 FLEX and ABL90 FLEX PLUS analyzers due to a risk of potentially inaccurate results. Owners must update the instructions for use to address changes in sample types, reportable ranges, configuration, and limitations.
How to tell if you have it
- Brand: Radiometer
- Product: ABL90 FLEX Analyzer
- Product: ABL90 FLEX PLUS Analyzer
- UDI-DI: 05700693930923, 05700693930909
- Software versions: 3.6 and lower
- ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O
- ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L
- Units affected: 5,395
The hazard
The in vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use. These changes include that sample types, including whole blood samples, are no longer suitable for testing FHbF and ctBil. The recall also addresses reportable ranges for some parameters, analyzer configuration, and limitations and warnings.
Context
GovInfoNow's records show Radiometer Medical ApS has 17 other recalls on record since February 27, 2018. Other recent recalls of similar products include Cepheid Xpert BCR-ABL Ultra Test Kits (October 30, 2025), Fujirebio Lumipulse G pTau 217 Plasma Calibrators (December 11, 2025), and Siemens Atellica CH A1c_E Hemoglobin Test (February 23, 2026).