The FDA announced on May 11, 2026, that Terex Utilities is recalling Power Over Fiber Base Modules because they lack required laser safety certifications and proper labeling. The agency mandates the recall of these devices distributed nationwide and internationally.
How to tell if you have it
- Brand: Terex Utilities
- Product: Power Over Fiber Base Module
- Part Numbers: P/N 633232, P/N 622365, P/N 633383, P/N 633910, P/N 635731
- Quantity: 101 units
The hazard
The devices were distributed without a laser product report filed with the CDRH and incorrectly labeled as Class IV instead of Class 1. They also omitted required labels under 21 CFR 1040.10 / IEC 60825-1, lacked manufacturer ID labels, and were not accompanied by compliant User Information.
Context
GovInfoNow's records show other recent recalls of similar products include the Repligen Maverick Measurement Module Spectrometer Recall (2025-12-10), the Beaver-Visitec OME6000U Laser System Recall (2025-12-05), and the Lutronic DermaV Laser System Recall (2026-08-06).