The FDA announced on June 24, 2026, that Medtronic Perfusion Systems is recalling EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR units due to potential pinhole leaks. The agency said the hazard may include hypovolemia, vessel laceration, or organ dysfunction during cardiopulmonary bypass.
How to tell if you have it
- Brand: Medtronic Perfusion Systems
- Product: EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR
- Model Number: 77422
- UDI-DI: 20763000135632, Lot Numbers: 0232446429 0232446444 0232486752 0232601515 0232601516 0232601522 0232601586 0232601591 0232601593 0232602180 0232798265 0232798353 0232849583 0232894981 0232895082 0232902694 0232902697 0232902698 0232902700 0232902803 0232902805 0232902846 0232985469 0233009922 0233108251 0233116991 0233127352 0233334923 0233334966 0233426694 0233426695 0233528727
- UDI-DI: 00763000135638, Lot Numbers: 0232446444 0232798353 0232895082 0232902697 0232902803 0232902846 0232985469 0233127352 0233334923
- Units affected: 21,290
The hazard
According to the FDA, pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Context
GovInfoNow's records show Medtronic Perfusion Systems has 31 other recalls on record since 2017-06-22. Other recent recalls of similar products include Medtronic Bio-Medicus Adult Cannula Kit Recall (2026-05-25), Medtronic Octopus Evolution Tissue Stabilizer Model TS2000 Recall (2026-05-12), and Medtronic DLP Cannula Recall (2026-04-21). Other recent recalls of similar products include Medline Convenience Kits Recall (2026-05-28) and Medline Convenience Kits with Swan-Ganz Catheters Recalled Due to Leak Risk (2026-05-28).