DxC 500i Clinical Analyzer Modules May Show Incorrect Test Results Due to Software Issue

Beckman Coulter recalls DxC 500i Clinical Analyzer Modules with specific part numbers due to a software issue that can cause test results to show falsely elevated values when plasma samples are tested.

🇺🇸 FDA recall #Z-0083-2027Medical deviceBeckman Coulter ›
✎ GovInfoNow plain-language explainer

Beckman Coulter is recalling DxC 500i Clinical Analyzer Modules that have a software issue. This issue can cause test results to show falsely elevated values when plasma samples are tested, potentially leading to incorrect medical diagnoses. Read more at govinfonow.com/recall/1629442-beckman-coulter-inc-recalls-dxc-500i-clinical-analyzer-modules-1.

Source Summary by GovInfoNow · govinfonow.com/recall/1629442-beckman-coulter-inc-recalls-dxc-500i-clinical-analyzer-modules-1 · Updated October 21, 2026 · FDA enforcement record: FDA #Z-0083-2027

Cite this page

“DxC 500i Clinical Analyzer Modules May Show Incorrect Test Results Due to Software Issue.” GovInfoNow, updated October 21, 2026. https://govinfonow.com/recall/1629442-beckman-coulter-inc-recalls-dxc-500i-clinical-analyzer-modules-1. Based on FDA recall #Z-0083-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0083-2027%22&limit=1).

The FDA announced on Sept. 2, 2026, that Beckman Coulter Inc. is recalling DxC 500i Clinical Analyzer Modules due to a software issue that may cause falsely elevated reported results. The agency classified the recall as Class II.

How to tell if you have it

  • DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522
  • DxC 500 AU Module, DxC 500i, Part No. C63521
  • Access 2 Module, DxC 500i, Part No. C13252; Software version 1.3.5 or earlier
  • UDI-DI 14987666545089
  • UDI-DI 14987666545072
  • UDI-DI 15099590742331

The hazard

According to the FDA, a software defect may cause the analyzer to revert test units to system defaults without notifying the user when testing plasma samples. This issue can result in falsely elevated reported results.

Context

GovInfoNow's records show Beckman Coulter, Inc. has 133 other recalls on record since June 20, 2017. Other recent recalls of similar products include Beckman Coulter DxC 500i Modules with Software 1.5 Risk: Erroneous Test Results (July 15, 2026), Beckman Coulter MicroScan LabPro V5.03 Panel Update 11 DVD Kit: Incorrect Algorithms May Affect Test Results (June 26, 2026), and Beckman Coulter Access Total T4 Calibrator Recall: False Low Results Risk Incorrect Treatment (March 16, 2026). Other recent recalls of similar products include Beckman Coulter REMISOL Advance Software Issue: Incorrect Test Results (April 29, 2026), Beckman Coulter DxC 700 AU Recall: Delay in Results May Cause Diagnosis Delays (Jan. 30, 2026), and AU480 Chemistry Analyzer May Show Error Message Causing Delay in Patient Results (Aug. 3, 2026).

How to tell if you have this one

Check for part numbers C63522, C63521, or C13252 with software version 1.3.5 or earlier. The recall includes specific UDI-DI numbers: 14987666545089, 14987666545072, and 15099590742331.

  • Brand
    Beckman Coulter
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide - US Nationwide distribution.
Recall numberProduct, codes & lots
Z-0083-2027DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252; Software version 1.3.5 or earlier. 1. UDI-DI 14987666545089 2. UDI-DI 14987666545072 3. UDI-DI 15099590742331
Z-0084-2027DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520; Software version 1.4.5 or earlier. 1. UDI-DI 14987666545058 2. UDI-DI 14987666545065
Based on the official FDA remedy

What to do — step by step

  1. Check Module Part Number

    Identify the part number on your DxC 500i Clinical Analyzer Module to see if it matches C63522, C63521, or C13252.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide - US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 42 days after the firm started it.

1. Firm started the recall: Sep 2, 202612. FDA classified it Class II: Oct 2, 202623. Published in FDA’s Enforcement Report: Oct 14, 202634. Added to GovInfoNow: Oct 21, 202642026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. Added to GovInfoNow

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0083-2027 (September 2, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0083-2027 to see it in context.
  • Plain-language summary written by qwen3:4b on October 21, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA